Presentation is loading. Please wait.

Presentation is loading. Please wait.

Itemiser 2 Maintenance Course Introduction.  Trace Detection  ITEMISER 2 Overview  Log-on and Calibration  Shift Maintenance  Weekly Maintenance.

Similar presentations


Presentation on theme: "Itemiser 2 Maintenance Course Introduction.  Trace Detection  ITEMISER 2 Overview  Log-on and Calibration  Shift Maintenance  Weekly Maintenance."— Presentation transcript:

1 Itemiser 2 Maintenance Course Introduction

2  Trace Detection  ITEMISER 2 Overview  Log-on and Calibration  Shift Maintenance  Weekly Maintenance  Monthly Maintenance  Questions and Comments Objectives

3 Trace Detection  Traces are microscopic amounts of particles and vapor  Traces are transferred to objects and people  The Itemiser 2 can detect traces to a billionth of a gram, or, a grain of salt cut into 1000 pieces

4 Trace Detection Sampling Itemiser 2 Sampling Demonstration

5 Trace Detection  What are explosive traces?  How are explosive traces transferred?  How can you describe a billionth of a gram?  How do you know when the Itemiser 2 detects explosive traces? Review

6 Itemiser 2 Overview Sample Trap Holder Display Screen Desorber Printer

7 Itemiser 2 Overview Select Scan Display Screen Status Indicator Software Version List of Substances Function Buttons

8 Itemiser 2 Overview Plasmagram Display Screen Plasmagram Substance List

9 Itemiser 2 Overview  Which Itemiser 2 component heats the sampling trap?  Which display screen highlights a detected substance in red?  What do the colors of the status indicator mean?  Which display screen shows a snapshot of the last sample?  How can you tell which version of software you are running? Review

10 Log-on and Calibration Log-on Demonstration

11 Log-on and Calibration  Calibration introduces a known substance into the system.  After the Itemiser 2 identifies the substance, the software adjusts the positions for all the other programmed substances.  Calibration can be affected by changes in temperature, atmospheric pressure, and humidity. What is Calibration?

12 Log-on and Calibration Calibration Demonstration

13 Log-on and Calibration  What are the different log-on numbers and their levels?  Why do you calibrate the Itemiser 2 ?  Which traps do you use to calibrate the Itemiser 2 ?  The calibration peak is normally found between which two numbers?  How often should you calibrate the Itemiser 2 ? Review

14 Shift Maintenance  Calibrate  Clean Fan Filter  Check Printer Paper  Check Supplies  Decontaminate Work Area  Record keeping

15 Shift Maintenance Shift Maintenance Demonstration

16 Weekly Maintenance  Perform Shift Maintenance  Power Down Unit  Clean Desorber  Inspect Filter  Clean Sample Air Line  Clean Membrane  Calibrate  Record keeping

17 Weekly Maintenance Weekly Maintenance Demonstration

18 Monthly Maintenance  Power Down Unit  Perform Weekly Maintenance  Change Dryer  Inspect Dopant  Change Membrane  Calibrate  Record keeping

19 Monthly Maintenance Monthly Maintenance Demonstration

20  Trace Detection  ITEMISER 2 Overview  Log-on and Calibration  Shift Maintenance  Weekly Maintenance  Monthly Maintenance  Questions and Comments Course Review

21 Detector Schematic


Download ppt "Itemiser 2 Maintenance Course Introduction.  Trace Detection  ITEMISER 2 Overview  Log-on and Calibration  Shift Maintenance  Weekly Maintenance."

Similar presentations


Ads by Google