APPLICATION TO THE HADROTHERAPY FOR OCULAR MELANOMAS G.A. Pablo Cirrone Qualified Medical Physicist and PhD Student University of Catania and Laboratori Nazionali del Sud INFN, Italy
2
LNS Superconducting Cyclotron in Catania is the unique machine in in Italy and South Europe used for protontherapy
3
Depth doses in radiotherapy: single fields
4
Why clinical proton beams? penetration depth is well-defined and adjustable most energy at end-of -range protons travel in straight lines dose to normal tissue minimised no dose beyond target
5
MEDICAL APPLICATION OF THE CATANA FACILITY INFN - LNS Choroidal Melanoma 85 - 90 % successfully treating with proton beams About 300 new cases for year in Italy In the world 5000 patients have been already treated with proton beams Age related subfoveal macular degenerations with subretinal neovascular membrane This is the main cause of third age blindness in EC countries It is a “social disease” for EC
6
0 ° respect the switching magnet 80 meter after extraction 3 m proton beam line LAYOUT OF LNS PRESENT TREATMENT ROOM
7
Scattering system Modulator & Range shifter Monitor chambers Ligth field Laser
8
Patient ready for treatment
9
DETECTORS USED FOR DOSE DISTRIBUTION MEASUREMENTS DEPTH DOSE DISTRIBUTION Markus Ionization chamber Silicon diode 2 mm Sensible Volume = 0.05 cm 3 LATERAL DOSE DISTRIBUTION GAF Chromic Film Thermoluminescent dosimeters Risolution 100 m for DDP and 200 m for LDP Markus Chamber layoutIrradiated GAF Chromic
10
‘PURE’ BRAGG PEAK EXPERIMENTAL RESULTS
11
MODULATED BRAGG PEAK
12
EXPERIMENTAL LATERAL DOSE DISTRIBUTION
13
Geant4 applications in prothon therapy our goals Simulation, design and realization of critical elements of a proton beam lineSimulation, design and realization of critical elements of a proton beam line Scattering systemScattering system Modulation systemModulation system Collimators along the beam lineCollimators along the beam line Three dimensions dose distribution verification in water at isocenter in the treatment conditionsThree dimensions dose distribution verification in water at isocenter in the treatment conditions Comparison with experimental and treatment planning data