PL-ELS 2100 Ice
2
PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination with the ‘evaporation gas’ technology to provide sub-ambient operation The ‘evaporation gas’ is increased to compensate for the increase in relative humidity at sub-ambient temperatures This approaches enables the PL-ELS 2100 Ice to operate at sub-ambient temperatures and as low as 10°C.
3
Even at ambient temperature the response of semi-volatile, low-molecular weight analytes can be poor. So the PL-ELS 2100 Ice has been designed to operate at sub- ambient temperature to improve the detection of semi-volatile compounds Introduction
4
PL-ELS 2100 Ice Design Features Nebulisation Same nebuliser as PL-ELS 2100 Temperature Range: 25-90°C Evaporation Integrated Cooled Evaporator Tube Temperature range: 10-80°C Active Heating/Cooling Same evaporation gas technology as PL-ELS 2100
5
Peltier Cooled Evaporator Tube
6
PL-ELS 2100 Ice Design Feature for Sub-Ambient Operation Aqueous eluents can be removed at 15°C Organic solvents (e.g. THF) can be evaporated as low as 10°C The benefit of operating below room temperature provides maximum sensitivity for low-molecular weight semi-volatile compounds.
7
PL-ELS 2100 Ice Design Feature for Sub-Ambient Operation The PL-ELS 2100 Ice can heat or cool rapidly, which reduces equilibration time. 35°C to 10°C in 10 minutes 15°C to 80°C in 11 minutes Accurate temperature control of evaporator temperature eliminates external temperature effects
8
PL-ELS 2100 Ice Design Features Detection Same Optical Bench as PL-ELS 2100 Digital Output 24bit (10Hz) via RS232 Faster Time Constant (0.1 Secs) for UPLC
9
PL-ELS 2100 Ice FOR VOLATILE SOLUTES Sub-ambient Evaporator temperatures required - 10-30°C – PL-ELS 2100 Ice Gas flow optimised for sub-ambient temperatures and mobile phase – 100% Water @15°C requires 2.2SLM of gas Compounds with high vapour pressures, low MW pharmaceuticals, pesticides, hydrocarbons
10
Sub-Ambient Applications The S/N ratio for thermally sensitive compounds such as Acetanilide is improved at sub-ambient temperatures In addition, the response of compound’s with high boiling points and high volatility, such as Alkanes, is dramatically improved. Hence, ELSD temperatures need to be as low as possible, to minimise volatility effects.
11
Sub-Ambient Applications: Acetanilide (MW 135)
12
Sub-Ambient Applications: Alkanes
13
Additional Benefits of the PL-ELS 2100 Ice Sub-ambient operation down to 10°C for maximum sensitivity to volatile compounds Active heating/cooling provides rapid equilibration between analyses. Improved accuracy of quantification.
14
Limit of Detection- PEG 106 x40 increase in S/N
15
What are you Missing?-Wine Analysis
16
PL-ELS 2100 Ice ‘Lock out’ Specs Sub-ambient operation down to 10°C High sensitivity to semi-volatile compounds DMSO transparency at ambient temperature Extremely low dispersion for high resolution separations Compatible with eluent flow rates, up to 5.0ml/min Rapid thermal equilibration Easy to use Extremely small footprint, stackable
17
ELS2100Ice is the best choice for ELS detection for volatile and semi-volatile compounds. ELS2100Ice has the widest range of sample detection for any ELSD, therefore it is the most ‘universal’ ELS Detector ELS2100Ice is the only ELSD with operation at sub-ambient temperatures. ELS2100Ice sub ambient specification can be used for lock- out specs vs competitive ELSDetectors. Key Points vs Competition
18
PL-ELS 2100/2100 Ice Specifications 200mm 450mm 200mm 415mm 450mm Light source: LED 480nm Detector: Photomultiplier tube Nebuliser temperature: ambient to 90°C in 1°C increments Evaporator temperature: PL-ELS 2100 Ice - 10 to 80°C (1°C increments) PL-ELS 2100 - ambient to 120°C (1°C increments) Gas flow rate: up to 3.25 SLM Gas pressure: typical 60-100psi, maximum 100psi Eluent flow rate: 0.2 to 5 ml/min Output: Analogue: 0-1V FSD Digital: 24bit, 10Hz using serial connection Communications: serial I/O (RS232), contact closure, TTL Power requirements: 90/120V AC or 220/250V AC 50/60 Hz Dimensions: 200x450x415mm (wxdxh) Part No.PL0860-0110 (110v), PL0860-0240 (240v)