1
Piezoluminescence:experimental data analysis Maciej Grzesiak, Jadwiga Jeziorska, Tomasz Ostafin Opole University Supervisor: mgr Grzegorz Bujnarowski
2
Purpose of the exercise Analysis of mechanical stresses with method of the spectroscopy in tester of ruby subjected to irradiating with bundle of heavy ions of the radiation in the function of the dose absorbed and in the function of the depth.
3
Piezoluminescence Changes in the luminescent radiation triggered with mechanical stresses. It manifesting through the appearance in the spectra of additional lines or strips (with additional centres of the luminescence e.g. vacancies ). These stresses influence to existing spectral lines which can cause shift them.
4
TimeOfFlight Cyclotron 670 MeVExperiment Bombed sample Measuring the profile of the bundle
5
Spectograph Shamrock SR - 303i. Shamrock SR – 303i parameters. Focal lenght (mm)303 Aperturef/4 Raciprocal dispersion ( nm/mm, nominal) 2.6 Mechanical scan range (nm)0 to 1450 Wavelenght resolution ( nm)0.1 Wavelenght accuracy (nm)+/- 0.2 Wavelenght reproducibility (nm)+/- 0.05 Focal plane size ( mm, WxH)28 x 14 Stray light ( meansured AT 20 nm from 633 nm laser line) 1.5 x10 -4 Experimental parameters of Shamrock SR – 303i filter – no filter beam energy – 670 MeV pattern’s temperature – 77,35 K angle beetwen pattern and beam - 60 degrees work temperature of CCD’s camera - - 60 degrees.(C)
6
CCD camera Grating Motorised Slits Cover Plates: used to control the effective height of the entrance slit (various sizes available) FLOW DIAGRAM.
7
Principle of confocal laser scanning microscope A laser beam passes a light source aperture, then is focused by an objective lens into a small focal volume within a fluorescent speciment. A beam splitter separated the light mixture by allowing only the laser light to pass through and reflecting the fluorescent light intothe detection appararus. After passing a aperture the fluorescent light is detected
8
Curve of the temperature relation for tester of ruby which for calibrating of curves served for two temperatures it is of liquefied nitrogen (77.35 K) and for room temperature (293 K). Spectre of tester of aroused ruby with laser lighting for two exchanged temperatures.
9
Model spectrum at the temperature of 77,35 degrees (Kelvin). As you can see maxsimum of the R2 peak is about 14460,19 cm -1, which is correct with results shown for the temperature’s model spectrum. R2 peak max The graph describing the change of the amplitude and shift spectral lines triggered with rise of heavy ion’s dose.
10
R2 peak R2 peak is about 14442,943 cm-1, which is correct with results shown for the temperature’s model spectrum. Model spectrum in the room temperature The graph describing the change of the amplitude depending on depth
11
This graph is presenting a dependence of the tension on an accepted dose of radiation. On this slide we can see a change of putting the maximum of the R2 depending on the depth
12
Thanks for coordinators of Bogoliubov - Infeld programme,Mr R. Zawodny and Mr W. Chmielowski and for the entire UC board of directors and of course our supervisor Mr. G. Bujnarowski for spending a beatifull moments in Dubna.
13
Thanks for attention