Download presentation
Presentation is loading. Please wait.
Published byEvangeline Shelton Modified over 9 years ago
1
NA62 Gigatracker Working Group Meeting 23 March 2010 Massimiliano Fiorini CERN
2
Measurement setup (1) M. Noy (19/11/2009) Discriminated pixel output
3
Measurement setup (2) single PC controlling everything via USB and Ethernet inject charge pulses with fixed rise time (2.5 ns) and variable amplitude Tektronics AFG3252 pulse generator driving charge injection circuit programmable 0 – 5 V, pulse min. 50 mV, steps of 1 mV 1/10 attenuation: 5 mV 0.5 V, steps of 100 µV charge injection ~20 fF (0.1 fC 10 fC) record injected signal and discriminated pixel output with oscilloscope LeCroy WavePro 7100A oscilloscope (20 GS/s for 2 channels, 1 GHz analogue bandwidth) CERN prototype chip: measurement of 1 of the 5 test pixels with discriminated output with CLOCK OFF with CLOCK ON (replaced USB card with ALTERA board) 250 MHz DLL operation (125 ps bin)
4
Measurement setup (3)
5
T in 0.035 V (0.7 fC) threshold
6
Measurement setup (3) T1T1 T2T2 0 V threshold
7
Injected pulse signals nominal pulse amplitudes from 0.06 V to 0.5 V (correspond to 1.2 fC and 10.0 fC resp.) note: 1/10 attenuation threshold set to 0.035 V (0.7 fC)
8
Discriminated output signals
9
Injection time measurement (1) least square method applied to: Method 1 3 samples above and below threshold Method 2 40 samples on the rising slopes at fixed position (~11- 13 ns)
10
Injection time measurement (2) least square method applied to: Method 1 3 samples above and below threshold Method 2 40 samples on the rising slopes at fixed position (~11- 13 ns) T in
11
Discr. times measurement least square method applied to: Method 1 3 samples above and below threshold (0 V) T1T1 T2T2
12
Correlation coefficient: T 1 and T 2
13
Correlation coefficient: T in
14
Ex. of low charge pulse shape (1)
15
Ex. of low charge pulse shape (2)
16
Method 1 Vs 2 for T in Measurement
17
T 1 jitter
18
Correction of Injection Baseline Variation
19
Injection baseline variation
20
T 1 jitter (w. baseline correction) variation of 0.035 V (0.7 fC) threshold taking into account (trace by trace) baseline variation lower T 1 jitter
21
Results with CLOCK OFF
22
Leading edge time (T 1 )
23
Trailing edge time (T 2 )
24
Injection time (T in )
25
(T 1 – T in )
26
(T 2 – T in )
27
Time over Threshold (T 2 – T 1 )
28
T 1 jitter: (T 1 – T in ) rms
29
T 2 jitter: (T 2 – T in ) rms
30
Results with CLOCK ON
31
Leading edge time (T 1 )
32
Trailing edge time (T 2 )
33
Injection time (T in )
34
(T 1 – T in )
35
(T 2 – T in )
36
Time over Threshold (T 2 – T 1 )
37
T 1 jitter: (T 1 – T in ) rms
38
T 2 jitter: (T 2 – T in ) rms
39
Extraction of CLOCK contribution
40
T 1 jitter: CLOCK contribution mean jitter value of (36 ± 2) ps above 2 fC
41
Geant 4 Simulation
42
Energy release GTK per hit mean energy: 72.4 keV (~20.1 k e-h ~3.2 fC) most probable energy: 53.7 keV (~14.9 k e-h ~2.4 fC) FWHM: ~25 keV (~6.9 k e-h ~1.1 fC) minimum energy: ~29 keV (~8.1 k e-h ~1.3 fC)
43
Charge-weighted T 1 jitter value taking into account the energy distribution of particle hits in the Gigatracker, one can extract a weighted average value for the jitter on T 1 no charge sharing between pixels considered
44
Charge weighted average result: (65 ± 12) ps CLOCK OFF Charge weighted average result: (76 ± 11) ps CLOCK ON CAVEAT measurements done with: 20 pF pixel input capacitance (no bump-bonded sensor) 2.5 ns pulse injection rise time 25 °C ambient temperature no charge sharing Comparison with simulation: simulations from J. Kaplon show 30-40 ps rms (160 ps pk- pk) for a 3.0 fC signal Results of weighted jitter
45
Conclusions T 1 jitter was measured to be: lower than 100 ps for charge injection greater than 2.0 fC ~69 ps at 3.2 fC injection (mean charge released by mip) Charge weighted jitter is (76 ± 11) ps with CLOCK ON Measurement + analysis method limited by noise especially for low charge injections (less than ~2 fC) measured jitter values have to be considered upper limits
46
SPARES
48
Energy release GTK per hit (2) mean energy: 72.4 keV (~20.1 k e-h ~3.2 fC) most probable energy: 53.7 keV (~14.9 k e-h ~2.4 fC) FWHM: ~25 keV (~6.9 k e-h ~1.1 fC) minimum energy: ~29 keV (~8.1 k e-h ~1.3 fC)
49
Time Over Threshold Correction
50
ToT Correction (1) build Look-Up Table to correct T 1 as a function of the signal Time Over Threshold, i.e. (T 2 – T 1 )
51
ToT Correction (2) from ToT plot fit, extract input charge values using 50 ps wide time intervals (from 10 ns to 20 ns)
52
ToT Correction (3) from (T 1 – T in ) plot fit, compute the time correction to be applied to T 1 then build complete LUT to be used in reconstruction program
53
Results (1) T 1 – T in distribution becomes flat error bar = jitter small shift around ~6 fC due to change in measurement conditions
54
Results (2) T1 jitter consistent with value before correction problem with the same 2 points around 6 fC
55
Results (3) problem with 2 points around ~6 fC due to events “leak” from one bin to the next one around the “threshold” for anomalous T in behavior
56
Results (3) problem with 2 points around ~6 fC due to events “leak” from one bin to the next one around the “threshold” for anomalous T in behavior
57
problem with 2 points around 6 fC due to events “leak” from one bin to the next one around the “threshold” for anomalous T in behavior Results (3)
58
Correction of Injected Pulse Height Variation
59
Inject. pulse height variation (1)
60
Inject. pulse height variation (2)
61
Inject. pulse height variation (3)
62
T 1 jitter selection of injected pulse height within ±0.1 mV of local mean value (±0.002 fC) comparable T 1 jitter (lower jitter for first bin only) cuts a lot of statistics and has negligible effect: not used in the analysis
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.