BLOOD PRESSURE BIOMEDICAL ENGINEERING 2006200401 TAE_EUI, KIM
2
Invasive BP measurement Catheter Dome Diaphragm Strain gages (usually Four) Pressure Sensor Po should be same as Pi “Fluid Mechanics”
3
What affect the quality of Po Catheter length should be shorter but there is Viscosity of liquid Diameter Elastic material for catheter Diaphragm itself should be elastic No air bubble
4
Equivalent Circuit Method of Catheter – Sensor System Electric CircuitFluid Mechanics Voltage, V [v]Pressure, P [pa] Current, i [A](Volume) Flow, f [m3/s] Charge, q [c]Volume, V [m3] R, L, C circuitPhysics things
5
Equivalent Circuit Method of Catheter – Resistance Electrical resistance : Liquid resistance : If viscosity is too big?
6
Equivalent Circuit Method of Catheter – Capacitance or Compliance ( AC flow ) The bigger Compliance, the bigger flow
7
Equivalent Circuit Method of Catheter – Inductance or Innertance
8
Equivalent Circuit Model Catheter Sensor
9
Equivalent Circuit Model Catheter & Sensor are rigid but not diaphragm
10
Equivalent Circuit Model
11
Frequency Transfer Function
12
Magnitude Response
13
Phase Response
14
Reference
15
Steady State Frequency Response
16
; Principle of Superposition
17
Transient Response ; Principle of Superposition Transient Property! Transient Property!