1 BLOOD PRESSURE BIOMEDICAL ENGINEERING 2006200401 TAE_EUI, KIM

2 Invasive BP measurement Catheter Dome Diaphragm Strain gages (usually Four) Pressure Sensor Po should be same as Pi “Fluid Mechanics”

3 What affect the quality of Po Catheter length should be shorter but there is Viscosity of liquid Diameter Elastic material for catheter Diaphragm itself should be elastic No air bubble

4 Equivalent Circuit Method of Catheter – Sensor System Electric CircuitFluid Mechanics Voltage, V [v]Pressure, P [pa] Current, i [A](Volume) Flow, f [m3/s] Charge, q [c]Volume, V [m3] R, L, C circuitPhysics things

5 Equivalent Circuit Method of Catheter – Resistance Electrical resistance : Liquid resistance : If viscosity is too big?

6 Equivalent Circuit Method of Catheter – Capacitance or Compliance ( AC flow ) The bigger Compliance, the bigger flow

7 Equivalent Circuit Method of Catheter – Inductance or Innertance

8 Equivalent Circuit Model Catheter Sensor

9 Equivalent Circuit Model Catheter & Sensor are rigid but not diaphragm

10 Equivalent Circuit Model

11 Frequency Transfer Function

12 Magnitude Response

13 Phase Response

14 Reference

15 Steady State Frequency Response

16 ; Principle of Superposition

17 Transient Response ; Principle of Superposition Transient Property! Transient Property!