1 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Chapter 14 Respiratory Monitoring in the Intensive Care Unit
2
2 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Learning Objectives After reading this chapter you will be able to: Identify the methods, normal values, and significance of measuring the following lung volumes in the ICU: tidal volume, rapid-shallow breathing index, vital capacity, functional residual capacity
3
3 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Learning Objectives (cont’d) Identify the methods, normal values, and significance of measuring the following airway pressures or related indices in the ICU: peak pressure, plateau pressure, compliance, airway resistance, mean airway pressure, maximum inspiratory pressure List the definition, methods of detection, and methods of minimizing auto-PEEP
4
4 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Learning Objectives (cont’d) Describe value of monitoring pressure, volume and flow waveforms, and pressure volume curves in mechanically ventilated patients Describe methods and significance of measuring the fraction of inspired oxygen and exhaled carbon dioxide in the ICU List the components of oxygen transport and their significance List components of clinical evaluation of oxygenation and their significance
5
5 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Learning Objectives (cont’d) Explain how these parameters can evaluate tissue oxygen delivery and use: Oxygen delivery and availability Oxygen consumption Mixed venous oxygen tension Venous saturation Arterial to mixed venous oxygen content difference Oxygen extraction ratio Blood lactate
6
6 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Learning Objectives (cont’d) Describe the value and limitations of pulse oximetry in monitoring oxygenation and oxygen delivery Identify the techniques for monitoring tissue oxygenation and utilization
7
7 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Overview Monitoring: repeated or continuous observations or measurements of the patient Guide therapeutic interventions Assess interventions Alert clinicians to changes in patient’s condition
8
8 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Ventilatory Assessment PaCO 2 : standard for assessing ventilation Changes in metabolism, lung mechanics, ventilatory efficiency, equipment function may precede changes in blood gas Ventilatory parameters monitored Lung volumes and flows Airway pressures Fractional gas concentrations
9
9 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
10
10 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Lung Volumes and Flows Why? They affect gas exchange They reflect changes in patient’s clinical status They indicate response to therapy They signal problems with patient/ventilator interface Who? Intubated and nonintubated patients
11
11 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Lung Volumes and Flows (cont’d) What do we measure? V T (5-8 ml/kg IBW) VT <5 ml/kg may indicate respiratory problem Pneumonia, COPD, CHF, ARDS, CNS depression Large VT in metabolic acidosis, sepsis, neurological injury VT = V A + V D V D = 25% to 40% of the VT V D >60% = need ventilatory support
12
12 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Lung Volumes and Flows (cont’d) High VT ventilation with positive pressure: volutrauma PEEP + smaller VT maintains FRC Discrepancies between set/measured VT Compressible volume of the circuit Inspiratory and expiratory flow sensors Pneumothorax Leaks in the circuit or ETT
13
13 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Lung Volumes and Flows (cont’d) Spontaneous breathing trial (SBT) failure: VT <300 ml or <4 ml/kg SpO 2 <85% to 90% Blood pressure and heart rate change >20% Respiratory rate >35/min Change in mental status Accessory muscle use Diaphoresis
14
14 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Rapid shallow breathing index (RSBI) RSBI = f (breaths/min)/VT (liters) RSBI >105: prognostic of failure V E = 5 to 6 L/min V E > 10 L/min: weaning not likely successful Lung Volumes and Flows (cont’d)
15
15 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Vital capacity (VC) 65 to 75 ml/kg IBW FVC <20 ml/kg preoperative: risk of pulmonary complications VC 10 to 15 ml/kg needed for deep breathing and coughing VC >10 to 15 ml/kg for successful weaning and extubation Lung Volumes and Flows (cont’d)
16
16 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Functional residual capacity (FRC) FRC = 40 ml/kg IBW PEEP and CPAP increase FRC Beneficial in atelectasis and refractory hypoxemia as occurs with ARDS Lung Volumes and Flows (cont’d)
17
17 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures Important to monitor airway pressures: Need for mechanical ventilation and readiness for weaning Determine site and cause of impedance to mechanical ventilation Evaluate elastic recoil, compliance of thorax Estimate amount of airway pressure transmitted to heart and major vessels Assess respiratory muscle strength
18
18 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures (cont’d) Peak pressure or PIP Pressure required to overcome opposition to airflow in the lungs Increased resistance Bronchospasm, airway secretions, mucus plugging Decreased compliance (lung or chest wall) Patient-ventilator interface problem
19
19 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
20
20 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures (cont’d) Plateau pressure Elastic recoil of lung and chest wall Static pressure during period of no gas flow Pressure required to maintain inflation
21
21 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
22
22 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures (cont’d) Mean airway pressure (Paw) Affected by CPAP, PEEP, inspiratory time (T I ), V T, PIP, and rate = Paw = [½ (PIP – PEEP) × (inspiratory time/total cycle time)] + PEEP Impacts oxygenation Caution when Paw >20 cm H 2 O _ _ _
23
23 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures (cont’d) Maximum inspiratory pressure (P Imax ) Influenced by: Respiratory muscle strength Patient effort/ventilatory drive Lung volume Phrenic nerve function Nutritional status Oxygenation/acid-base status Normal P Imax : –80 to –100 cm H 2 O P Imax > –30 cm H 2 O may be useful to predict successful weaning
24
24 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures (cont’d) Auto-PEEP Total PEEP – set PEEP Reduction of auto-PEEP Bronchodilator therapy Decreased TI (allows more time for exhalation) Reduction of mechanical frequency Reduction of V E is the most effective
25
25 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures (cont’d) Compliance Dynamic compliance = {Corr VT – [(PIP – EEP) × CF]}/PIP – EEP Static compliance = Corr V T - [(P plat - EEP) × CF]/P plat – EEP Normal static compliance in patients receiving mechanical ventilation: 40 to 80 ml/cm H 2 O Compliance <20 – 25 ml/cm H 2 O associated with failure to wean
26
26 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Airway Pressures (cont’d) Airway resistance (Raw) (Ra-w = PIP (cm H 2 O) – plateau pressure (cm H 2 O)/flow (L/sec) Normal = 1 to 3 cm H 2 O/L/sec _ _
27
27 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Integrating Pressure, Flow, and Volume Evaluating patient/ventilator interface At the patient: use of accessory muscles, color, diaphoresis, heart rate, respiratory rate At the airway: type, size, integrity, stability At the ventilator circuit: leaks, temperature, condensate At the ventilator settings and monitoring panel
28
28 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Integrating Pressure, Flow, and Volume (cont’d) Monitoring pressure, flow, and volume Graphic display screen Scalar: a single parameter over time Pressure-time waveform Volume-time waveform Flow-time waveform Loop: two parameters in a continuous tracing Pressure-volume loop Flow-volume loop
29
29 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Integrating Pressure, Flow, and Volume (cont’d) Titrating PEEP and tidal volume with P/V Static pressure-volume curve Super syringe technique Time consuming and cumbersome Useful in acute lung injury Lower inflection point + 2 cm H 2 O: minimal PEEP Upper inflection point: overdistention
30
30 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
31
31 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Fractional Gas Concentrations FIO 2 Exhaled CO 2 Capnometry and capnography Affected by temperature changes, shivering, seizures, trauma, high carbohydrate infusion Measure efficiency of ventilation PETCO 2 : accurate estimate of PaCO 2 CPR effectiveness
32
32 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
33
33 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Dead Space-Tidal Volume Ratio V D /V T Anatomic dead space: 1 ml/kg IBW Alveolar dead space V D /V T = Pa CO 2 – PE- CO 2 /Pa CO 2 Normal: 25% to 40% _
34
34 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Evaluation of Oxygenation Evaluation of oxygen transport Oxygen consumption Oxygen delivery (DO 2 ) Oxygen reserves Oxygen content CaO 2 = (Hb × 1.34 × % saturation) + (PaO 2 × 0.003) Cardiac output Oxyhemoglobin dissociation curve
35
35 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
36
36 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
37
37 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Evaluation of Oxygenation (cont’d) Monitoring adequacy of arterial oxygenation Partial pressure of arterial oxygen Should be kept 60 to 80 mm Hg Alveolar-arterial oxygen tension difference P(A-a)O 2 PaO 2 /FIO 2 ratio ALI: 200 to 300 ARDS: <200
38
38 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Monitoring adequacy of arterial oxygenation Oxygen index = Paw × F IO 2 × 100/Pa O 2 OI >40: mortality >80% __ Evaluation of Oxygenation (cont’d)
39
39 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Intrapulmonary shunt (QS/QT) Increased in atelectasis, pneumonia, ARDS, pulmonary edema Pulse oximetry will reveal low Sp O 2 with elevated F IO 2 Co-oximeter useful to determine Ca O 2.. Evaluation of Oxygenation (cont’d)
40
40 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Monitoring Tissue Oxygenation Oxygen delivery (DO 2 ) Cardiac output × CaO 2 × 10* Normal: 550 to 650 ml/min/m 2 Oxygen consumption (V O 2 ) Fick principle Normal: 100 to 140 ml/min/m 2 *Change vol% to ml/L.
41
41 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
42
42 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Monitoring Tissue Oxygenation (cont’d) Mixed venous oxygen tension Pv O 2 Normal: 38 to 42 mm Hg Low: inadequate cardiac output, anemia, hypoxia High: poor sampling technique, left-to-right shunt, septic shock, increased cardiac output, cyanide poisoning _
43
43 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
44
44 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Monitoring Tissue Oxygenation (cont’d) Mixed venous oxygen saturation (Sv O 2 ) Fiberoptic reflectance oximetry Decreased: suctioning, shivering, extubation, weaning, PPV _
45
45 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc.
46
46 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Monitoring Tissue Oxygenation (cont’d) Arterial-mixed venous oxygen content difference, C(a – v) O 2 reflects: Normal: 4 to 6 vol% Increased: low cardiac output, increasing V O 2 Decreased: septic shock, increased cardiac output, anemia, left shift ODC _.
47
47 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Monitoring Tissue Oxygenation (cont’d) Oxygen extraction ratio [C(a – v) O 2 /Ca O 2 ] Normal: 25% to 30% Increased: low cardiac output, increased VO 2, decreased CaO 2, Decreased: high cardiac output, sepsis _.
48
48 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Monitoring Tissue Oxygenation (cont’d) Blood lactate Anaerobic metabolism = lactic acid production Normal: <1.7 to 2.0 mM/L >3.83 mM/L = 67% mortality >8 mM/L = 90% mortality
49
49 Mosby items and derived items © 2010 by Mosby, Inc., an affiliate of Elsevier Inc. Summary Common ventilatory measurements Lung volumes and flows Airway pressures Fractional gas concentrations These measurements allow clinicians to determine the need for mechanical ventilation, monitor the patient, and determine the readiness for weaning