LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 LIGO S2 Calibration Michael Landry LIGO Hanford Observatory Representing the calibration team: Rana Adhikari, Peter Fritschel, Gabriela Gonzalez, Luca Matone, Brian O’Reilly, Hugh Radkins, Patrick Sutton, Akiteru Takamori
2
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Current status Online calibration with accuracy of ~20% was available with 1 hour latency Interim calibration issued with corrected DC calibration Final calibration to include hardware measurements: accurate representation of the actuation function Additional Wednesday talks relating to calibration include: Brian O’Reilly, Patrick Sutton, Xavier Siemens, and Soumya Mohanty
3
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Calibration Basics I R(f): Response Function Scale Factor (from Calibration Line) G(f): Open Loop Gain = C(f) D(f) A(f) X i / AS_Q(f i,t) = R(f i,t) = [1 + (t) (t) G(f i ) ] / (t) C(f i ) X i = AS_Q(f i,t). R(f i,t)
4
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Calibration Basics II We choose a time t 0 and measure 0,G 0 and R 0. This allows us to extrapolate to any time t using the amplitude of the calibration line (or lines). The amplitude (t) is a function of the alignment, dependent on the carrier power in the arms and the sideband power in the recycling cavity. (t) is tracked throughout the science segments. R(f i,t) = [ 1 + (t) (t) G(f i ) ] / (t) C(f i )
5
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 DC calibrations Have redundant measurements, including sign toggling, fringe counting, free- swinging and PZT fine actuator Gained confidence with multiple methods during S2; excellent agreement Will limit the number of redundant measurements during the S3 run; making some of these beforehand Concentrate on reducing error
6
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 L1 anti-imaging filter
7
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 New model comparison
8
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 LHO open loop gains
9
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Propagating the calibration: LLO
10
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Propagating the calibration: LHO H1
11
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Contributions to the Error Magnitude: Phase: R(f i,t) = 1 + (t) (t) G(f i ) (t) C(f i )
12
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Estimating errors in L1 ’s
13
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Violins in H1 Independent check by Rachel Berkowitz (SURF), and Fred Raab – modelling of violin modes in H1
14
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 S2 strains
15
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 LLO epoch comparsion
16
LIGO-G030423-00-DM. Landry – Hanover LSC Meeting, August 18, 2003 Conclusions Necessary to incorporate more realistic filters in model, i.e. measurements instead of as-designed filter Final calibration with errors to be issued by next Monday Expectations for S3: same 1hr latency for online calibration, with accuracy as per existing (final) calibration – we learn more each time