Presentation is loading. Please wait.

Presentation is loading. Please wait.

A method of step height and plate thickness measurement within the unambiguous range of two laser wavelength interferometer Naoyui Koyama.

Similar presentations


Presentation on theme: "A method of step height and plate thickness measurement within the unambiguous range of two laser wavelength interferometer Naoyui Koyama."— Presentation transcript:

1 A method of step height and plate thickness measurement within the unambiguous range of two laser wavelength interferometer Naoyui Koyama

2 Michelson interferometer Use two lasers simultaneously Laser a wavelength : λa Laser b wavelength : λb Laser a and laser b do not interfere each other

3 Example 1 λa/λb =11/13 λaλa λbλb λa+λb Moire pattern Intensity variation within “beat” B is inverse of “beat” A Intensity varies periodically A → B → A → B

4 Example 2 ABA Intensity variation within “beat” B is inverse of “beat” A λa =635nm λb =785nm

5 → Modifications of Michelson interferometer DETECTOR : single detector → CMOS(or CCD) detector MIRROR : perpendicular to the reference beam → slightly inclined No part of the instrument is mechanically movable, making possible precise measurements free of backlash.

6 Optical path length of reference beam of column r is w(s-r)(sin2θ ) longer than column s CMOS detector

7 Step height is (1/2)(s-r)w(sin2θ) CMOS detector

8 How to measure thickness of a plate No plate is placed in the pass of measurement beam, or infinitesimal thin plate is set at zero point of optical path length difference, both case get same intensity variation as Fig. 1.

9 How to measure thickness of a plate Intensity variation is periodic, when infinitesimal thin plate is placed at, same intensity variation as placed at is seen (Fig. 1).

10 How to measure thickness of a plate Measurement beam passed through the prism is reflected by the plate. When the plate is placed appropriately, intensity variation is the same as Fig. 1. Intensity variation of measurement beam not pass through the prism is shifted as seen in Fig. 2. Thickness of the plate d is calculable from this shift.

11 An example instrument to measure intensity variation and corresponding optical path length difference.

12 How optical axes of two wavelength laser diode coincide with each other by prism. Compensation plate

13 Two beams partially overlap each other. Two wavelength Michelson interferometer is usable as one wavelength Michelson interferometer simultaneously.

14 References N. Koyama, A method of step height measurement within the unambiguous range of two laser wavelengths interferometer, Optik, 126 (2015) 313-316. All slides shown here can be seen in my website. N. Koyama, http://www12.plala.or.jp/sokkyo/prism.html