Download presentation
Presentation is loading. Please wait.
Published byBritton Fox Modified over 9 years ago
1
Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton, Kavita Gala ASCRS 2010 No Relevant Financial Disclosure
2
Purpose To measure the reproducibility of optical measurements in pseudophakic patients using a double pass measurement system. Light scatter Increasing interest in measurements of light scatter and impact on visual function Methods of measuring scatter Double-pass Psychophysical Hartmann-Shack
3
Variables that affect light scatter Tear film Cornea Anterior Chamber Intra-ocular lens Vitreous clarity Measurement error inherent in the system may also affect the repeatability of measurements Montes -Mico et al IOVS 46:12:4468-44742005 Nanavaty et al Ophthalmogica in press
4
Optical Quality Analysis System II (OQAS) Uses double-pass measurement system Images point spread function Derives MTF, Strehl ratio, light scatter (objective scatter index - OSI) (Visionmetrics, Spain) MTF - the ratio between contrast in the retinal image and contrast of the original object. OSI - objective parameter calculated by evaluating the amount of light on the outside of the double-pass image in relation to the amount of light on the centre. Higher OSI value relates to more intraocular scattering. Strehl ratio - measurement of an optical system (ratio of the MTF to that of a diffraction limited system). Thus, Strehl ratio will be a figure between 0 and 1, the higher its value, the better
5
Methods Prospective reproducibility study Population Normal pseudophakic eyes, At least three months post-cataract surgery Measures Refraction OQAS measurements used customised OQAS II machine
6
Methods Refraction - to obtain the cylindrical correction to optimise measurement. OQAS II - Pupils were dilated and measurements made for a 4mm exit pupil diameter. 3 sets of measurements of OSI and Strehl ratio from a single eye at one visit were taken and the average calculated. This was repeated one to two weeks later. Correlation plots and Bland-Altman analyses were used to assess and compare the 2 sets of measurements.
7
Results 51 eyes from 29 patients were assessed, with a heterogenous group of intraocular lenses: 20 eyes with an Alcon SA60AT 5 eyes with an Akreos MI60, 6 eyes with an Acrismart 36a IOL, 6 eyes with an Alcon SA60WF 6 eyes with the Morcher 89A ‘Bag in the Lens’ 2 eyes with Crystalens HD intra-ocular lens. 6 eyes with unknown IOL type
8
Results R 2 = 0.78 Average of 2 nd reading Average of 1 st reading Correlation of Strehl ratio values
9
Results Correlation of OSI values R 2 = 0.95 Average of 1 st reading Average of 2 nd reading
10
Results Bland-Altman Plot of Strehl ratio Coefficient of Repeatability = 0.033 Strehl ratio Difference Standard Deviation = 0.017
11
Results Bland-Altman Plot of OSI Difference Ocular Scatter Index Coefficient of Repeatability = 1.56 Standard Deviation = 0.79
12
Conclusion Optical measurements using a double-pass system are reproducible despites concerns of changes in the anterior segment between visits.
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.