Download presentation
Presentation is loading. Please wait.
Published byAdele Riley Modified over 9 years ago
1
12/12/09 MPGD 研究会 @ 神戸大学 1 Yusuke Komatsu A B. Azmoun B, C. Woody B, K. Ozawa A University of Tokyo A,Brook Haven National Lab. B
2
Outline HBD for PHENIX Results from the last year run Gain - gas fraction & V GEM dependence (ArCF4) Gain - Time dependence with Tech-Etch GEM Comparison between Tech Etch and CERN Time dependence as a function of the amount of water 2
3
Hadron Blind Detector for PHENIX Windowless Cherenkov light detector Dalitz rejector for measurements of low–mass e + e - pairs.(m e+ e- ≤ 1 GeV) Dalitz rejection with opening angle under “no” magnetic field Improve S/B ratio ~2 orders(goal) in mass spectrum! Signal: ex) φ → e + e - BG: π 0 → e + e - γ γ → e + e - arXiv:0706.3034 PHENIX 200GeV Au+Au Run4(before install HBD) 3
4
HBD Concept Mesh GEM CsI Readout pad Reverse bias e-(+) Cherenkov light Charged particle(γ<γ th ) Forward bias pe e- CF4 4 CsI photocathode is coated on the top GEM. CF4 both for Cherenkov radiator and electron multiplication gas. (as an electron multiplication) CF4 : UV E cutoff =11.5eV → λ=108nm + CsI : limit from QE → λ=200nm Sensitive region of wave lengh is 108< λ[nm] <200 ~40 photoelectrons /e -(+)
5
View & performance of HBD Run9 5 cm 55 cm e-e- e+e+ Pair Opening Angle Triple GEM detectors (10 panels per side) Gas & HV 22 pe few pe Hadron blind Single electron signal
6
Installed HBD for Run10. 6
7
Objective ◎ Base measurements of GEM for further upgrades Improved gas mixture Ar & CF4 have all most 100% transmittance in the sensitive wave lengh region. Ar mixed gas can reduce the operation voltage. -gain measurement in ArCF4 7 Green Ar Red CF4
8
Objective Time dependence of gain and H2O contamination Used Tech-Etch GEM. They are said to have different property from CERN GEM in time dependence of gain. And H2O ppm effects gain. -gain vs time (Tech-Etch) -gain vs time changing H2O ppm (Tech-Etch) 8
9
Measurements of gain vs V GEM in ArCF 4 Measured triple GEM gain vs V across GEM(V GEM ) with Fe 55. Dependence of the ArCF4 ratio. Mesh GEM Readout Pad DG4.09 mm, 1 kV/cm TG 1 1.59 mm, 2.5 kV/cm IG 1.57 mm, 3 kV/cmTG 2 1.61 mm, 3.5 kV/cm GEM made by CERN Pitch : 120μm Hole size(outer): 80μm (Inner ): 50μm V GEM 9 Fe 55
10
GEM CERN-segmented 10 Each resistor is 10MΩ. One side is composed of 4 segments.
11
GEM chamber 11 Electrodes for GEM H2O: <100ppm
12
ArCF 4 gain vs V GEM 12 20/80 Green line represents the effective voltage for mixed gas to obtain the same signal strength as pure CF4. Results show ArCF4 has enough gain.
13
Energy Res. of ArCO 2 and ArCF 4 13
14
Gain vs time measurement: ArCO 2 70/30 (Tech-Etch) Measured time dependency of gain in ArCO2 70/30. Used another GEM series. They were made by Tech- Etch. Insulator CERN: apical TE:E-type Kapton Pitch : 120μm Hole size(outer): 80μm (Inner ): 50μm Double conical gap[mm]electric field[kV/cm] DG4.111 TG11.672.5 TG21.683 IG1.653.5 14
15
Relative gain 15 Gain of Tech-Etch GEM increased by ~3.5 times in 2.5 hrs.
16
Gain vs V GEM in plateau Gain was extremely high compared to CERN-segmented(gain abs.=1342@340V, ArCO2 70/30). 16
17
Gain curve :TE changing H 2 O & CERN 17 TE ~50 ppm 36%increase TE ~250ppm 31.1% TE ~1100ppm 1100ppm 38.1% C-C-TE 330~140ppm 23% C-C-C 330~130 ppm 28.4% When reset charge up, flow Ar+H2O at ~2 l/min for 10 min. Tech-Etch GEM need longer time to reach gain saturation than CERN GEM. H2O can decrease a saturation time,however could not see such effects.
18
Summary Measured gain vs V GEM in ArCF4. Energy resolution did not depend on gain but a kind of gas mixture. Gain changed according to time ~3.5times eventually when used Tech-Etch GEM. About Tech-Etch GEM, water content in the chamber changed the gain, but did not the time to reach a saturated point of gain. 18
19
Backup 19
20
Requirements for HBD Very high e -(+) efficiency Double hit resolution at least 90% when opening angle is small. Hadron blindness and high sesitivity to a small number of photoelectrons as signal. ~10 4 multiplication factor Use GEM and CsI photocathode. Use analog information about charge. Need photo electrons as many as possible. Respond to a large bandwidth of Cherenkov light. pure CF4 bandwidth:6~11.5eV 20
21
GEM chamber (closed) 21
22
ArCO 2 gain vs V GEM 60/40 22
23
Gain value ArCO2 Ar_CO2 Vgem[V]90_1080_2070_30(abs gain)60_4050_5040_60 270450 280945 2901972 3004312649 31095932613 32080935116 33010462 340217951342379 350164022676641 36049621136263 37097711955418 380187283571670 390323521105294 400172671857439 410660 4201016 23
24
Ar_CF4 Vgem[V]80_2070_3060_4050_5040_6030_7020_8015_85 350416 360756364 3701455658 38027381159472 39053962125826 40039321447556 4102539965403 42046791673679 43028981159460 44050831993745 45033751280553 4602076894634 47034691479931 48024841504 49040942519 5004034 510 520 Vgem[V]10_90 475709 4851130 4951776 5053051 Gain value ArCF4 24
25
Results: ArCO 2 Energy resolution 25 Consistent with a reference thesys NIM A523(2004)345-354. ~22%.
26
Results: ArCF 4 Energy resolution 26
27
ArCO 2 90/10 CERN-segmented Left)V GEM =320V, Right)310V 27
28
28 ArCO 2 CERN-segmented Left)70/30 400V, Right)80/20 400V
29
Gain was saturated after 45hours. 29
30
Gain curve changing H 2 O ppm (TE GEM) 30 38.1% increase 31.1% Tech-Etch GEM need longer time to reach gain saturation than CERN GEM. H2O can decrease a saturation time,however could not see such effects. 36.0% ~1100 ppm ~250ppm ~50 ppm
31
QE of CsI photocathode 31 From Craig’s slide.
32
But GEM had tripped on the way. “11/12 9:36 all GEM tripped.Ramped up 350V again. There seems no damage.” Kept on measurement. Gain still increased. 32
33
Results: Time dependency of gain ArCO2 70/30,V GEM =350V 33
34
34
35
35 I.Ravinovich Electron pairs – Central Arms (200 GeV pp)
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.