Fort Sill - POOF Tim Onasch, Mike Timko March 18, 2007 (day 0) LOG Day 1 – co-adds 2, used 1 multiply data by 2 Day 2 – IE used 34, should be 15, multiply data by 4/10
2
Fort Sill POF Summary Day 0 – setup March 18, 2007 (Sunday) Day 1 – AM Paladin 155mm howitzer (NO offline for part of the day) (Mon) Day 2 – PM small arms fire – SAWs and machine gun (PTR down for final hour) (Tues) Day 3 – AM 105 mm howitzer, PM single shot grenade launcher M203 (Wednesday) Day 4 – AM/PM single shot SAW’s (AMS off line for the entire measurement) (Thursday)
3
Post Fort Sill To-do list Is 10-sec GA working properly? 3/23 looks OK (look at final hdf files from 3/22/07 to confirm) How is the noise level? At ARI talk to JJ and AT to diagnose noise problems – do we need to open things up? Clean instrument and sort through cables Why is heater bias not heater controlled? Get latest software –Igor, Squirrel, and DAQ Standard cold start procedure –Turn on pumps (wait 10 min) turn on oven, bake overnight –Turn filament and let heat up for a couple of hours –Boot up TPS controller with no_amp_noK settings –Baseline in bitwise –m/z cal –Tune including heater bias (maximize 28, max res of 32/184, optimize peak shape, minimize 39/23) –Redo baseline/ m/z cal –Get SI in bitwise (SI was about 18 on last days of POF) –BFSP calibration in GA mode (IE_cal menu, 60 sec) with Ben’s software (IE was about 5e-7 on last days of POF) –Take 10 min of 10-sec GA truck and 10-sec GA on filter – analyze with Squirrel – Are things OK?
4
070318_poof_day0
6
Valve open, chopper open
7
Valve closed, chopper open
8
Valve open, chopper closed
10
10-sec scan routine
14
0.3 sec wait time
15
Day 1: M155 Range ? March 19, 2007
18
Shot 2 Mass Specs
19
Ambient ptof before shot 1
20
shot 1 size distribution
21
Ambient ptof
22
shot 6 size distribution
23
Combined data from Day 1 morning - Paladin
24
Day 2 March 19, 2007
25
3-sec ptof truck air
26
1-sec ms truck air
27
3-sec ms truck air
28
This single ion was done post-fact, the data from day 2 needs to be multiplied by 4/10
29
Bfsp_ie/day2bBFSP 070320 afternoon, AB = 6e5, IE = 1e-6
30
Day 3 105 in the morning Grenade launchers in the afternoon? 3/21/07
31
Stable SI
32
Small correction to m/z cal
33
BFSP ie cal afternoon of Day3 070321, IE = 5.4e-7, AB = 3.5e5
34
Loads of K signal at m/z = 39 Also, really running our mcp into the ground (possibly due to K) Raise board gain from 0.2 to 0.1 Lower MCP Tune away from K at m/z = 39 Here are the results
40
This is a screen shot from the end of day 3 Truck air
41
This is a screen shot from the end of day 3 Truck air
42
Day 4 prep
43
Single ion went up a very slightly, probably due to MCP and thresholding
44
Tuning looks good, single extraction is reasonable,
45
Morning/Afternoon Day 4 March 23, 2007 Massive failure – no MCP signal even for tuning purposes Why? A) about 1 hr of oven temp > 900C b) very bumpy ride to the range Diagnosis? None, lots of external connections checked out OK, waited for several hours and the signal was recovered. At first, there was nasty electronic noise, but that went away after about an hour. What happened? Unknown at this time. The following slides are the tuning and recals performed the afternoon of Day 4
56
1 sec ms truck air
57
1 sec ms filter air
58
1 sec ptof truck air
59
1 sec ptof filter air
60
30 sec GA truck air