1
An alternative spectrograph mount Bruce C. Bigelow University of Michigan Department of Physics 5/14/04
2
2 Spectrograph mount Objectives: Design a high-performance spectrograph mount to FP high stiffness / high first resonance low launch stresses, small deflections simple, clear interface to focal plane (ICD) ease installation and removal of spectro. minimize mass, part count Performance goals (from Besuner, 4/29/04) first resonance above 100 Hz material failure (yield) above ~40g acceptable focal plane distortion for D T = 160K
3
3 Spectrograph mount Design features: Truss structure, Invar tubes 2:1 Truss geometry Horizontal translations, no tilts Attaches to common focal plane mounting points Essentially no loads carried by focal plane assembly Simple interface to spectrograph 3 discrete support points, or round flange Supports spectrograph load near center of mass minimizes moment loads Simple interface to FP (mount points, cylindrical volumes) Spectrograph and mount easily separate from FP or each other
4
4 Spectro mount on FP 2:1 truss geometry – inner to outer triangle dimensions
5
5 Spectro mount on FP FP assembly with spectro structure attached
6
6 Spectro installed Ease of access to detector connections FP assembly with spectrograph included (note redundant str.)
7
7 Spectrograph mount FE Analyses: General Spectro mass modeled as 10Kg Invar cylinder Invar spectrograph mounting truss 7075 Alum. focal plane baseplate (extreme case) Static Strut geometry: 25 mm OD, 2 mm wall thickness 40 G transverse acceleration (launch orientation) Dynamic First 10 mode shapes, frequencies Thermal Delta T = -160K Stresses and deflections
8
8 Static FEA Deflections: Ay = 40G, Dy = 137 microns (0.005”)
9
9 Static FEA Stress: Ay = 40G, s = 23.7 MPa (3436 Psi) (250 MPa yield)
10
10 Dynamic FEA First resonance = 413 Hz, transverse mode First 6 freq: 1.413 Hz 2.415 Hz 3.416 Hz 4.470 Hz 5.478 Hz 6.490 Hz
11
11 First resonance (.avi)
12
12 Thermal FEA Elements: Purple – Al 7075, Red, Blue - Invar
13
13 Thermal FEA Axial deflection for -160 K = 940 microns (0.037”), due to Al baseplate (d l = 50 microns for 0.25m of Invar). Much less for MZT baseplate, can be thermally compensated with strut materials.
14
14 Thermal FEA Stress for -160 K = 67.4 MPa (9773 Psi) (worst case with Al baseplate, simple model)
15
15 Spectrograph mount Conclusions: Truss structure meets all design objectives: high stiffness, high resonance low stresses simple, clear interface to FP easy installation, removal, and access to FP minimum part count, low mass Truss structure meets all performance goals: first resonance above 400 Hz (100 Hz goal) material failure safety factor of 10 at 40g minimal or no distortion of focal plane (independent) Truss is a viable alternative to flexure mount