New Flowcell – How Easy? 1
2
2
3
SenCell 3
4
Small, robust, easy to handle Stepless variable spacing 4
5
SenCell - Specifications Working electrode GC 2mm Reference electrode Salt Bridge - blue ISAAC - green HYREF – black 5
6
Sensitivity – wall jet Flow cell design A. thin-layer B. wall-jet C. porous flow-through 6
7
Cotrell Equation 7 ECD Cottrell equation i lim = 1.47 n F (F m ) 1/3 (DA/b) 2/3 C i lim = C spacer thickness A= WE area A b = spacer thickness C = concentration
8
Variable spacing – more signal Stepless variable spacing Smaller spacing Higher signal 8
9
Spacing 9
10
Sensitivity – more signal 100um vs 12 um spacing 10
11
Sensitivity – more signal Stepless variable spacing Smaller spacing Higher signal 11 0 5000 10000 15000 20000 020406080100 Spacing SenCell (µm) S/N of 100 nM standard DA 5-HT
12
Sensitivity LOD 50 pM DA (0.23 fmol) 65 pM 5-HT (0.25 fmol) 100 pmole/L standards in Hac + Ringer, 5 uL injection 12
13
Pressure Higher flow rate due to UHPLC Smaller spacing down to 10um 13
14
Sencell Max Pressure Leak tight ring Rating: 25 bar 14
15
UHPLC compatible High speed analysis Fast response Noise level time (min) 00.51.0 [nA] Current 0 5 10 15 DA 5-HT 100 nmole/L standards in Hac, 5 uL injection 15
16
UHPLC compatible time (min) 00.51.0 [nA] Current 0 5 10 15 DA 5-HT 100 nmole/L standards in Hac, 5 uL injection 16
17
Conclusion SenCell – The most Easy to use flowcell Wall jet – performance Variable spacing - more signal Less noise by improved fluid dynamics UHPLC compatible 17