Download presentation
Presentation is loading. Please wait.
Published byMalcolm Mitchell Modified over 9 years ago
1
6th Summer School on “ADC & DAC Metrology” PETRČESÁK Czech Technical University PhD Student Analysis of the system for dynamic ADC testing
2
Prague, Czech Republic
3
Czech Technical University
4
System for dynamic ADC testing
5
BAND-PASS FILTER
6
BAND-STOP FILTER
7
Distortion recalculated to the band-pass filter output (measured in dB) common low-cost filtershigh-quality filters 4.416 MHz 9.485 MHz 19.51 MHz 4.416 MHz 9.485 MHz 19.51 MHz THD–109–108–96< –130 SFDR11110997> 139
8
Frequency spectra of testing signals with frequency is 19.5 MHz SETUP: generator – band-pass filter – attenuator 6 dB – band-stop filter amplifier +25 dB – digitizer NI PXI-5122