Detection of Ziram by LC/APCI/MS Hiroshi MORIWAKI
Osaka City Institute of Public Health & Environmental Sciences
2
Ziram use toxicity pesticide antifouling agent endocrine disruptor
((CH3)2NCS.S)2Zn M.W m.p. 240 Log Pow 1.09 endocrine disruptor mutagen carcinogen
3
Analysis of Ziram × Conventional method
Direct detection by GC/MS → × Conventional method Decomposition of the ligand to CS2 × Chemical speciation of dithiocarbamate pesticides LC/MS ( Direct detection of ziram )
4
Ion mode Ionization Positive ion mode Negative ion mode ESI/MS × ×
APCI/MS × Corona current<0.6mA × ○ Corona current>0.6mA
5
Mass spectrum B) A)
6
Corona current vs. peak intensity
7
Cone voltage vs. peak intensity
8
Conditions of APCI/MS
9
Column separation
10
Mass chromatogram
11
Calibration curve
12
Summary Next step ・ Direct detection of Ziram by LC/APCI/MS
・ Corona current > 0.6mA ・ LC separation: GPC column ・ Instrumental detection limit: 2mg/L Next step ・ Development of the pretreatment ・ Use of LC/MS/MS ref. H. Moriwaki, Y. Yamaguchi and M. Fukushima: Rapid Commun. Mass Spectro., in press (2001).