Download presentation
Presentation is loading. Please wait.
Published byNancy Garrison Modified over 9 years ago
1
Engineering Perspective on Application of Simulated Ground Motions Jonathan P. Stewart & Emel Seyhan University of California, Los Angeles Robert W. Graves USGS Pasadena
2
Outline Need for verification, validation, and calibration (VVC) Validation procedures Validation of ShakeOut motions Calibration of high frequency IMs from Graves-Pitarka hybrid procedure
3
Need for VVC Verification: consistency of motions across computational platforms for common conditions Bielak et al., 2010 ShakeOut: example vel.(t)
4
Need for VVC Verification Validation: results of analysis generally consistent with recordings
5
Need for VVC Verification Validation Calibration: adjustment of parameters to achieve desired attribute of simulated motions
6
Validation Procedures Waveform comparison using earthquake data Graves and Pitarka, 2010
7
Validation Procedures Waveform comparison using earthquake data IM comparison to earthquake data Closest Distance (km) Graves and Pitarka, 2010
8
Validation Procedures Waveform comparison using earthquake data IM comparison to earthquake data IM comparison to GMPEs Frankel, 2009
9
ShakeOut Validation Motions from hybrid procedure
10
ShakeOut Validation Motions from hybrid procedure Analysis of residuals – 4 NGA GMPEs – R, V s30, Z for site i
11
ShakeOut Validation Motions from hybrid procedure Analysis of residuals Too-fast distance attenuation SO PH R R Star et al., 20xx
12
ShakeOut Validation Motions from hybrid procedure Analysis of residuals Too-fast distance attenuation Too-low intra-event standard deviation Star et al., 20xx
13
Calibration Adjustment to high frequency component of hybrid procedure
14
Calibration Adjustment to high frequency component of hybrid procedure Increase Q to slow distance attenuation Seyhan et al., 20xx
15
Calibration Adjustment to high frequency component of hybrid procedure Increase Q to slow distance attenuation Seyhan et al., 20xx
16
Calibration Adjustment to high frequency component of hybrid procedure Increase Q to slow distance attenuation Randomize Fourier amplitude spectra Seyhan et al., 20xx A = f(M)
17
Calibration Adjustment to high frequency component of hybrid procedure Increase Q to slow distance attenuation Randomize Fourier amplitude spectra Seyhan et al., 20xx
18
Conclusions Engineers want to use simulated motions Needed for conditions having sparse data Verification – Validation – Calibration – Each step documented – Avoid circularity in use of data
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.