Download presentation
Presentation is loading. Please wait.
Published byMatthew Shepherd Modified over 9 years ago
1
LED Source Tests RSL August 2012
2
http://www.oceanoptics.com/technical/LLSLightSource.pdf Blu Loop LLS Cool White – see the last page of: http://www.oceanoptics.com/Products/bluloop.asp t 600 um core fiber, goes to small integrating sphere; plan is to swap at source
3
fiber going to source(s); P/N QP- 600-2-SR integrating sphere; M/N FOIS-1 SEI M/N SR-3500 spectroradiometer SR-3500 input optical fiber zoomed sphere/SR-3500 fiber end (it is inside the sphere)
4
test plan Warm up: take a dark, turn on source, take one lite (an average of 10 scans) every 5 min for 3 – 4 hours Repeatability: repeat with the other LED source by changing the 600 um patch fiber at the source end (also covers warm-up) Do this for a couple of days (tests repeatability of source level and effect of decoupling and recoupling the SMA connector at the source/fiber junction) Turn on both sources, leave them on several days, swap fiber back and forth, acquiring data every 2 h or so Move SEI bare fiber to NPR to get a calibration If time and data look good, move sources to temperature/humidity chamber and measure levels
Similar presentations
© 2025 SlidePlayer.com. Inc.