Presentation is loading. Please wait.

Presentation is loading. Please wait.

Ventilator Graphics Emeritus Professor Georgia State University

Similar presentations


Presentation on theme: "Ventilator Graphics Emeritus Professor Georgia State University"— Presentation transcript:

1 Ventilator Graphics Emeritus Professor Georgia State University
Vijay Deshpande, MS, RRT, FAARC Emeritus Professor Georgia State University Atlanta, Georgia, USA Adjunct Visiting Professor, Manipal University College of Allied Health Sciences Manipal, Karnataka, INDIA

2 Mechanical Ventilation
SCALARS LOOPS Graphics

3 Inspiratory Flow Pattern
Peak inspiratory flow rate PIFR Beginning of expiration exhalation valve opens Inspiration Insp. time TI Expiratory Time TE Flow (L/min) Total cycle time TCT Time (sec) Beginning of inspiration exhalation valve closes Expiration

4 Expiratory Flow Pattern
Beginning of expiration exhalation valve opens Inspiration Expiratory time TE Time (sec) Flow (L/min) Duration of expiratory flow Expiration Peak Expiratory Flow Rate PEFR

5 Auto-PEEP Time (sec) Flow (L/min) Inspiration Normal Patient
} Air Trapping Auto-PEEP Expiration

6 Inadequate Inspiratory Flow
Active Inspiration or Asynchrony Patient’s effort Normal Abnormal Time (sec) Flow (L/min)

7 Volume Ventilation breaths are Volume Cycled
Inspiratory Tidal Volume Volume Cycling Volume (ml) Inspiration Expiration TI Time (sec)

8 Air Leak Volume (ml) Air Leak Time (sec)

9 Components of Inflation Pressure
PIP } Transairway Pressure (PTA) Exhalation Valve Opens Paw (cm H2O) Pplateau (Palveolar) Inspiratory Pause Expiration Begin Inspiration Time (sec) Begin Expiration

10 Inadequate Inspiratory Flow
Inadequate Flow Adequate Flow (cm H2O) Paw Time (sec)

11 Assisted Mode (Volume-Targeted Ventilation)
Patient triggered, Flow limited, Volume cycled Ventilation Flow L/m Pressure cm H2O Preset VT Volume Cycling Volume mL Time (sec) From: Essentials of Ventilator Graphics- An interactive CD. Vijay Deshpande, MS, RRT and Ruben Restrepo, MD, RRT. Available at

12 SIMV (Volume-Targeted Ventilation)
Flow L/m Pressure cm H2O Volume mL Spontaneous Breaths

13 SIMV + PS (Pressure-Targeted Ventilation)
Time-Cycled Flow-Cycled Pressure Flow Volume (L/min) (cm H2O) (ml) Set PC level Set PS level Time (sec) Fig 7.21 PS Breath

14 PSV Flow L/m Pressure cm H2O Volume mL Time (sec) Set PS level
Flow Cycling Set PS level Pressure cm H2O Volume mL Time (sec)

15 PSV Flow-Cycle Criteria
Peak Inspiratory Flow Flow 40% 20% 5% Simply put, the lower the ESENS, the longer the time before inspiration ends. The higher the ESENS setting, the shorter the inspiratory time. In general, the most appropriate ESENS setting is compatible with the patient’s condition, neither extending nor shortening the patient’s intrinsic inspiratory phase. ESENS can be adjusted by patient assessment. If inspiratory times appear to exceed the inspiratory efforts of the patient, a higher ESENS can be attempted to see if patient comfort improves. If the breath appears to be terminating prematurely, a lower ESENS can be set and patient comfort re-evaluated as well as it’s effect on tidal volume. T 36 36 36 36 36

16 CPAP + PSV ( BiPAP ) Flow Cycling Volume mL Flow L/m Pressure cm H2O
Set PS level Pressure cm H2O CPAP level Volume mL Time (sec) Fig 7.31

17 Components of Pressure-Volume Loop
VT Expiration Volume (mL) Inspiration PIP Paw (cm H2O)

18 Air Leak Volume (ml) Air Leak Pressure (cm H2O)

19 Inadequate Inspiratory Flow
Volume (ml) Active Inspiration Inappropriate Flow Normal Abnormal Paw (cm H2O)

20 Increased Raw Higher PTA Normal Slope Lower Slope Pressure (cm H2O)
Vol (mL) Normal Slope Lower Slope Pressure (cm H2O)

21 Lung Compliance Changes and the P-V Loop
Volume Targeted Ventilation Preset VT COMPLIANCE Increased Normal Decreased Volume (mL) Paw (cm H2O) PIP levels

22 Over-distention Preset Tidal Volume Volume (ml) Pressure (cm H2O)
With little or no change in VT Normal Abnormal Volume (ml) Paw rises Pressure (cm H2O)

23 Lung Compliance Changes and the P-V Loop
Volume Targeted Ventilation Preset VT COMPLIANCE Increased Normal Decreased Volume (mL) Paw (cm H2O) PIP levels

24 Flow-Volume Loop Inspiration PIFR FRC VT PEFR Expiration Flow (L/min)
Volume (ml) FRC VT Flow (L/min) PEFR Expiration

25 Air Leak Normal Abnormal Inspiration Expiration Volume (ml)
Flow (L/min) Volume (ml) Air Leak in mL Normal Abnormal Expiration

26 Air Trapping Normal Abnormal Inspiration Expiration Flow (L/min)
Does not return to baseline Volume (ml) Normal Abnormal Expiration

27 Assisted Mode (Pressure-Targeted Ventilation)
Patient Triggered, Pressure Limited, Time Cycled Ventilation Time-Cycled Pressure Flow Volume (L/min) (cm H2O) (ml) Set PC level Time (sec)

28 Scalars in Assisted Mode
FIGURE 1

29 Pressure Control Ventilation Flow – Time Scalar
TIME CYCLING TI Time Fig 8.16

30 PCV with long T I or Decreased Lung Compliance
Flow No Flow Period Exhalation Valve Opens Time Set TI Inspiratory flow Decreases to zero Fig 8.17

31 Pressure Control Ventilation with inadequate TI
Flow TIME CYCLING Exhalation valve opens TI Time Fig 8.18

32 Pressure Control Ventilation - Effect of Changes in Lung Compliance on
Flow-time scalar A : Decreased Compliance B : Normal compliance C : Increased Compliance Flow TIME CYCLING C A B TI Fig 8.19

33 Thank You !

34 Apnea Alarm Patient becomes Apneic Disconnection in Sedated Patient
Oversedation in Patients on Low rate SIMV or PCV

35 “Ventilator inoperative” Alarm
Absent Electrical Source (Unplugged) Power Failure “Ventilator inoperative” Alarm Internal Battery has very low charge

36 High Minute Volume or High Respiratory Frequency
Pain Anxiety High Minute Volume or High Respiratory Frequency Hypoxemia

37 Low PEEP Alarm Active Inspiration Inappropriate Alarm setting
A leak in the circuit

38 Other Useful Alarms High Minute Volume
Check Sensitivity setting for trigger level If using external nebulizer, reset the alarm until the treatment is completed, and then reset it to the initial setting. Check physiological causes of increased minute ventilation Low FiO2 Check Gas Source pressures Check Oxygen sensor for proper function (using oxygen analyzer)

39 Kinking or displacement of the tube
Decreased Lung Compliance Increased Resistance Secretions Bronchospasm Kinking or displacement of the tube Pneumothorax Pulmonary Edema, Pleural Effusion High Pressure Alarm Asynchrony (patient fighting) Inadequate Flow Auto PEEP, Improper Sensitivity Coughing

40 Low Pressure Alarm Leak In the System, Associated with Tracheal Tube
Cuff Leak, Chest tube leakage Associated with low minute ventilation or low tidal volume Low Pressure Alarm Patient Disconnection Proximal Pressure Line Disconnected

41 Low Tidal volume or Low Minute Volume
IN VOLUME VENTILATION IN PRESSURE VENTILATION Leak In the System Partial Disconnection Increase in f or decrease in Tidal Volume ( In Volume Ventilation) Increased Resistance and/or Decreased Lung compliance (Leak in the system does NOT Trigger low volume alarm)

42

43 Double triggering Volume-targeted ventilation. Notice the circled double volume breath and the pressure-time Waveform to drop below baseline triggering another breath immediately following the first breath. Hallmark of double triggering-peak pressure of the second breath rises as the second breath is stacked on the first breath. From Nilsestuen J, Hargett K. Using Ventilator Graphics to Identify Patient-Ventilator Asynchrony. Resp Care 2005;50(2):

44 Double triggering Causes
Commonly occurs with VC continuous mandatory ventilation Also with PSV with high termination flow criterion High respiratory drive Insufficient respiratory support such as minute ventilation or tidal volume too low with a high respiratory rate. Patient’s Itime (neural) is different (longer) than the ventilator set Itime Small Vt and short inspiratory time

45 Missed triggers Missed trigger with pressure control ventilation using the expiratory flow waveform. The arrow points to the positive flow deflection that is the hallmark sign of a missed trigger. From: Branson R, Blakeman T, Robinson B. Respiratory Care, 2013

46 Ineffective inspiratory efforts corrected by flow and expiratory time
COPD receiving assist-control ventilation with a constant Vt of 0.55 L. Arrows indicate ineffective ventilation with flow rate at 30 L/min in A. Black arrows indicate ineffective inspiratory efforts. Inspiratory flow rate was increased to 90 L/min which decreased inspiratory time and prolonged expiratory time. This reduced dynamic hyperinflation and the number of ineffective ventilations was reduced. Increasing flow rate can help reduce ineffective triggering in COPD patients with dynamic hyperinflation. During pressure control ventilation reducing inspiratory time will prolong expiratory time to help reduce dynamic hyperinflation. From:Kondili E, Prinianakis G, Georgopoulos D. Patient-Ventilator interaction. Br J Anaesth 2003;91(1):

47 Inspiratory Rise Time

48 Inspiratory Time Inspiratory Time
Too Short May be able to deliver more tidal volume if inspiratory time lengthen to baseline Normal: Good, Flow into lung has stopped, no extra pressure held in lung that is not wanted Too Long: No Flow at end inspiration Pressure is held in lung, may cause increase lung injury


Download ppt "Ventilator Graphics Emeritus Professor Georgia State University"

Similar presentations


Ads by Google