Presentation is loading. Please wait.

Presentation is loading. Please wait.

Douglas P. Moehs Emittance Probe 1 (side) Distance from slit to detector: 4.470 inches Detector with multiple collectors Expected collector width based.

Similar presentations


Presentation on theme: "Douglas P. Moehs Emittance Probe 1 (side) Distance from slit to detector: 4.470 inches Detector with multiple collectors Expected collector width based."— Presentation transcript:

1 Douglas P. Moehs Emittance Probe 1 (side) Distance from slit to detector: 4.470 inches Detector with multiple collectors Expected collector width based on individual measurements: 0.008 (8 mils) 1 copper strips and 1 Kapton strip make up a collector Width of 1 copper strip: 6 mils Width of Kapton: 2 mils Average Collector Width: 9 mils The detector is comprised of 22 collectors plus 1 additional Kapton strip The total detector width as measured: (aprox.) 0.201 The average collector width is defined as the detector width – the width of 1 Kapton strip/ number of collectors [(0.201-.002)/22] Angular Resolution: (2 mrad)  ~ Sin(  ) = ave. collector width / distance from slit Measured distance from shield to detector : (aprox.) 1.72 inches

2 Douglas P. Moehs Emittance Probe 2 (top) Distance from slit to detector: 4.525 inches Detector with multiple collectors Expected collector width based on individual measurements: 0.008 (8 mils) 1 copper strips and 1 Kapton strip make up a collector Width of 1 copper strip: 6 mils Width of Kapton: 2 mils Average Collector Width: 8.8 mil The detector is comprised of 22 collectors plus 1 additional Kapton strip The total detector width as measured: (aprox.) 0.196 The average collector width is defined as the detector width – the width of 1 Kapton strip/ number of collectors [(0.196-.002)/22] Angular Resolution: (1.95 mrad)  ~ Sin(  ) = ave. collector width / distance from slit Measured distance from shield to detector : 1.72 inches (from probe 1)

3 Douglas P. Moehs

4 Probe 1 and Probe 2 Output Channel mapping Conversion To current Channels 20 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Probe pin numbers 21 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Outer most Inner most Shield Pin 22 Pin 23,24 are not used Pin 25 grouned

5 Douglas P. Moehs Basic Operation Probes 1-4 all use the same amplifiers. Thus each time a probe is used the correct cable in the back of rack LE1-RR2-10 needs to be connected to the sample and hold unit. The ACNET channels are: HVEM00 through HVEM20 The motor control channels are: PROBE1 through PROBE4 Probes 1 and 2 cannot both be in the beam at the same time because they occupy the same space. Thus they are interlocked such that one must be in the “out” position before the other can be inserted. Similarly for probes 3 and 4. To run Probes 3 and 4 the safety system beam stops need to be out. This means the Linac needs to be secured. However you do not need to accelerate beam because these probes are at the front of Tank 1. A +700 volt bias on the shields seems to yield the “cleanest” signal however this means that secondary electrons dominate the signal. Beware! The production of secondary electrons on surfaces is very dependant on the incident angle of the primary ion. See the next slide for a rough correction.

6 Douglas P. Moehs Probe2 Secondary Electron Correction For additional information on secondary electron emission see Huashun Zhang, Ion Sources, p. 12- 13, 1999. The secondary electron yield (SEY) is the number of liberated electrons per incident particle, in this case negative hydrogen. SEY change nonlinearly with the incident particle energy and is dependant on the incident angle scaling roughly as the secθ of the the incident angle. The probable signal adjustment for Probe 2 is shown here. To correct the probe signal multiply by the values given here.

7 Douglas P. Moehs Probe 1 Secondary Electron Correction


Download ppt "Douglas P. Moehs Emittance Probe 1 (side) Distance from slit to detector: 4.470 inches Detector with multiple collectors Expected collector width based."

Similar presentations


Ads by Google