PR570 POWERITE II TRAINING Dedicated to Success in Electronic Servicing for 50 Years.

Similar presentations


Presentation on theme: "PR570 POWERITE II TRAINING Dedicated to Success in Electronic Servicing for 50 Years."— Presentation transcript:

1 PR570 POWERITE II TRAINING Dedicated to Success in Electronic Servicing for 50 Years

2 “POWERITE II” 6 WAY SAFETY ANALYZER 1. AC Receptacle Tester & Voltage Monitor 2. AC Variable Power Supply 3. Isolation Transformer 4. Output Current/Power Monitor 5. Output Circuit Breaker 6. Safety Leakage Tester

3 Measures Input ACV to Safety Ground Resets Isolated Output Trip Current Measures Variable ACV at Isolated Output Isolated Variable AC Output Select to View Trip Current Measures AC Leakage Adjusts Isolated Output Trip Current Level PR570 Equipment Familiarization

4 AC Electrical Outlet

5 AC Electric Box

6 CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT Hot To External Earth Ground Earth Ground

7 CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT AC Receptacle Tests

8 A Variable AC Transformer

9 Variable AC Power Supply 0-140 VAC CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT

10 Hot Ground Direct off-line DC supply with bridge rectifier results in a “HOT” chassis “HOT” ground has direct path to electrical hot side of electrical outlet Approximately 67 VDC between “HOT” ground and Earth ground Connecting a ground from a test instrument to a “HOT” ground causes high current through the diode

11 An isolation transformer electrically isolates the “HOT” side of the AC plug, eliminating the voltage difference and current that may result between ground references

12 Defeating Isolation The isolation transformer can be defeated if twohot chassis are connected at the same time.

13 Isolated Output On/Off Control CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT Isolated Output On Indicator (On when Lit.) Output Off - Output On Switch(Switches Isolated Output Voltage On or Off)

14 Isolated Output Current & Power Monitoring CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT Isolated Output Current & Watts Meter

15 Isolated Output Circuit Breaker CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT Position To Meter Circuit Breaker Level 0-4Amps Reset if Circuit Breaker Trips Control Sets Circuit Breaker 0-4A

16 If the “HOT” ground of the power supply becomes electronically connected to exterior parts of the cabinet housing, a shock hazard exists. Protect yourself, your business and your customer. ALWAYS perform a leakage current check after completing a repair!

17 Causes of Excess AC Leakage Current Defective AC Line or Switching Supply Isolation Transformers Leaky AC Bypass Filter Capacitors Improper Part Installation - Safety Critical Isolation Parts Foreign Objects - Coins, screws, paper clips

18 SMPS Safety Leakage Paths Switch Mode Transformer Primary to Secondary AC Input Filter Capacitors Isolation Filter Capacitors Between Hot & Isolated Grounds

19 Effects of AC Leakage

20 A safety leakage test ensures that the monitor doesn’t have AC line leakage to exposed metal parts Virtually all manufacturer’s service literature requires a safety leakage test be performed at the completion of any repair Safety leakage testing applies to all consumer electronic devices powered from the AC line, whether or not they have a 3-wire power cord Underwriter’s Laboratory (UL) standard is 500  a Other standards may apply International Electrotechnical Commission (IEC) standard is 700  a Safety Leakage Testing

21 Underwriter’s Laboratory (UL) standard is 500  a International Electrotechnical Commission (IEC) standard is 700  a 1Do not use an isolation transformer for this test. Plug the completely reassembled unit directly into the ac outlet. 2Connect a 1.5K, 10W resistor paralleled by an.015uF capacitor between each exposed metal cabinet part and a good earth ground such as a water pipe, as shown above. 3Use an ac voltmeter with at least 5000 ohms/volt sensitivity to measure the potential across the resistor. 4The potential at any point should not exceed 0.75 volts. A leakage current tester may be used to make this test; leakage current must not exceed 0.5 milliamps. If a measurement is outside of the specified limits, there is a possibility of shock hazard. The chassis should be repaired and rechecked before returning it to the customer. 5Repeat the above procedure with the ac plug reversed. (Note: An ac adapter is necessary when a polarized plug is used. Do not defeat the polarizing feature of the plug.) Manufacturer Safety Leakage Test

22 CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT Safety Leakage Testing

23 Checking Leakage Test Calibration CURRENT/POWER TRIP RESET CURRENT POWER LEAKAGE TRIP SET HOT CHASSIS Amps uAmps Watts uAmps VARIABLE ISOLATION TRANSFORMER & SAFETY ANALYZER PR570 MIN MAX MIN ONOFF ISOLATED OUTPUT VOLTAGE TRIP SET 100 uAMP LEAKAGE CAL CHECK Set Output to 120VAC & Touch Safety Probe Here. 3 Amps At 117VAC Continous, 4 Amps Intermittent Use TO EXTERNAL EARTH GND Refer To Pull Chart Isolated Output On When Lit AUTO-TOGGLE AC Line Safety Gnd Both Must Light SAFETY PROBE INPUT POWER POWERITE II WARNING: For Equipment Under Test Only! ! ISOLATED OUTPUT Set Isolated Output to 120VAC. Leakage Test Should Read 100uA +/- 2 Counts

24 Measures Input ACV to Safety Ground Resets Isolated Output Trip Current Measures Variable ACV at Isolated Output Isolated Variable AC Output Select to View Trip Current Measures AC Leakage Adjusts Isolated Output Trip Current Level PR570 Function Review

25 Congratulations! You are now an experienced PR570 User


Download ppt "PR570 POWERITE II TRAINING Dedicated to Success in Electronic Servicing for 50 Years."