1
Valery Ray Particle Beam Systems & Technology, Methuen, USA vray@partbeamsystech.com CAD – less Blind Navigation in Focused Ion Beam System PBS&T Chris Gerlinsky Titus Consulting, Saskatoon, Canada chris@nopnop.ca
2
1/5/2016 ISTFA 20052 Purpose Develop methodology for blind coordinate navigation on undocumented devices. Improve accuracy of navigation in standard FIB equipment. Ensure successful FIB contacting of small buried nodes.
3
1/5/2016 ISTFA 20053 Outline Methodology for acquisition of the device layout information. FIB coordinate navigation by acquired layout. Enhancements of navigational accuracy.
4
1/5/2016 ISTFA 20054 Study of the Device Layout Acquire high-resolution image Identify functional blocks Identify reference points Locate nodes of interest Layout of random off-the-shelf device
5
1/5/2016 ISTFA 20055 Identification of Reference Objects Stepper alignment mark reference identified using optical microscope. The same alignment mark is easily recognizable using FIB.
6
1/5/2016 ISTFA 20056 Enhanced Accuracy “Lock” in FIB Outline reference object by repair boxes; Mark center of the object with a cross; Make a precision lock on the center, even if the object is damaged by de-packaging or FIB milling;
7
1/5/2016 ISTFA 20057 Identification of Nodes and Acquisition of Preliminary Coordinates Analyze layout and identify nodes targeted for contacting; Trace nodes to locations convenient for FIB access; Acquire bitmap X/Y coordinates of the access locations and use as FIB coordinates; Layout of random off-the-shelf device
8
1/5/2016 ISTFA 20058 Acquisition of Precision Coordinates De-process location of the targeted node in a FIB and acquire precision coordinates of the node.
9
1/5/2016 ISTFA 20059 Enhanced Reliability of Contacting Contacting of buried node (green) by rectangular 0.2μm X 0.4μm access via provides safety margin to avoid shorting to the above- located nodes (yellow).
10
1/5/2016 ISTFA 200510 Enhanced Reliability of Contacting Monitoring of sample absorbed current provides reliable endpoint detection with high S/N ratio.
11
1/5/2016 ISTFA 200511 Summary Developed methodology for identification of IC node coordinates without manufacturer’s CAD data. Developed and used improved techniques of coordinate – based navigation in FIB system. Achieved reliable FIB contacting of deeply buried nodes by multiple 0.2µm X 0.4µm vias.