1 Electrode Array GK-12 Meeting November 2nd 2010
Megan Easterly

2 Electrochemistry Oxidation Reduction Loss of electrons
Electrode removes electrons from the solution Anodic peak current Reduction Gain of electrons Electrode gives electrons to the solution Cathodic peak current

3 Electrodes Working (WE) Reference (RE) Auxillary (AE) Electrolyte
Where the reaction of interest occurs Current is measured Reference (RE) Measures WE potential by maintaining a stable potential Auxillary (AE) Electron provider/remover Electrolyte Electrically conductive solution

4 Gold Electrode Array 8 Au working 1 Au reference 1 Au auxillary
2 3 4 5 6 7 8 aux ref 8 Au working 1 Au reference 1 Au auxillary Study FDCA

5 Ferrocenedicarboxylic Acid (FDCA)
E0 ferrocene = V vs. NHE 0.197 MORE positive Bard, A.; Faulkner, L., “Electrochemical Methods. Fundamentals and Applications” 2nd Ed. Wiley, New York. 2001 Image courtesy of Dr. Karst research group:

6 The Math Randles-Sevcik equation
is a collection of constants at 25 oC 0.443 * Fardays constant * universal gas constant Should be linear under a diffusion controlled case Bard, A.; Faulkner, L., “Electrochemical Methods. Fundamentals and Applications” 2nd Ed. Wiley, New York. 2001

7 Simultaneous Detection
External reference and auxillary

8 Simultaneous Detection
Normalized currents to electrode area Linearity > Sensitivity (μA*mm-2*mM-1) Ox 1.43 ± 0.07 Red ± 0.07

9 DPV of FDCA Optimal potential determination Ran individually
vs. AgAgCl 0.43 V vs. Au 0.37 V Ran individually

10 Amperometry of FDCA On-chip reference and auxillary
0.37 V Individual detection

11 Amperometry of FDCA Normalized to electrode area Linearity Sensitivity
> Sensitivity 22.44 ± 12.93 nA*mm-2*mM-1

12 New Mask Well defined areas Working = 0.09 mm2 Reference = 2.25 mm2
Auxillary = 2.25 mm2 Spacing = 0.3 mm

13 Future Work Fabrication of new electrode array
Chemiluminescence detection of NO donor Calibration of electrodes using NO Modification of Au array Nafion Polypyrrole Diab, N.; Oni, J.; Schulte, A.; Radtke, I.; Blochl, A.; Schuhmann, W., Talanta, 61, 2003, 43-51