ProStar 335 Detector Wednesday, 26 April 2017
Petro van Poppel – Field Support Specialist Wednesday, 26 April 2017
2
Overview Production Tests with 0x0 flow cell
Introduction to PS 335 Detector Dual Array option Poly Verify Software Flow cell Communications Poly Verify Production Tests with 0x0 flow cell Wavescan (Settings Tab) (Input/Output Tab) (EEPROM) (COMMS Tab) (Calibration Tab) PS 335 Shipping Test Problems
3
ProStar 335 Mechanical Optical layout
4
ProStar 335 Ray Trace Layout
5
335 Optics
6
335 Ray Trace
7
335 Single and Dual Array Single Array 1024 photodiode sample array 2 photodiode references (250nm and 500 nm) No extended range spectra Uses only 9x0 and 4x0 flow cells Microbore to Analytical applications Dual Array 1024 photodiode sample and reference array Extended range spectra (up to AU) Used for analytical to prep apps
8
Flow Cell Ray Beam General design of the 9x0 flow cell
9
Flow Cell Specs Flow Cell Prep (9 x 1) Super Prep (4 x 0.15)
Analytical (9 x 0) Micro (4 x 0 ) Pressure rating 1000 psi Flow Cell Volume 30 µl 145 µl 15 µl 1.5 µl Material (non optical) Ti & PEEK Noise Performance 200 ml/min 10-500ml 1x10-4AU 800 ml/min L 40 ml/min 0.1 µl-100 ml 5x10-3AU 1 ml/min 0-20 ml 7.5x10-3AU
10
Communications TCP/IP communications for Detector: Static or Dynamic
Networked or standalone Firewall Access on Network, OS or PC
11
PS 335 PolyVerify Software
12
PS 335 PolyDiagnostics Ctrl-Shift-F10 to enable Diagnostics side
Ctrl-Shift-F9 to hide Diagnostics side
13
335 PolyVerify Test Certificate
Test Certificate provided by the factory for each 335 detector
14
PS 335 PolyVerify Tests
15
PS 335 Noise Test
16
PS 335 Wavelength Accuracy test
Run with a 4 nm slit
17
PS 335 SBW Accuracy Test Wavelength Range of 190 to 950 nm
18
PS 335 Detector Linearity Test
pump head pressure of at least 100 atm using a back pressure restrictor of 40psi .
19
PS 335 Wavescan
20
PS 335 Wavescan Results This wavescan seems to have a rather large difference between the sample and reference paths (long and short paths). Possible reasons for this are bad calibration, dirty flow cell, air bubbles in flow cell, foggy optics.
21
PS 335 Contributions from each Lamp
22
PS 335 PolyDiagnostics Mode
23
PS 335 Input/Output Mode
24
PS 335 Main PWB Power Supply LED’s
C A: + 24 V power Supply B: + 9V and – 9V Power Supply C: + 5 V Power Supply Front of the Unit
25
PS 335 Identifying the Array Type
Knowing the Array type is necessary when running the Input/Output test diagnostics
26
PS 335 Dual Array P8 disconnect
P8 connection
27
PS 335 PolyDiagnostics EEPROM
28
PS 335 PolyDiagnostics COMMS
29
PS 335 PolyDiagnostics Calibration Tab
30
Caffeine Shipping Test
PS 335 Caffeine Shipping Test: Test made up of 1 components: Loop injection using 25 µg/ml caffeine Test is conducted using Galaxie Software
31
PS 325/335 Caffeine Loop Injection Test
Test Calculates: Peak Height Peak Area Noise S/N
32
Common Problems PS 335 Error Messages Table can be found: High Noise
Lamp on, not ready SIG_CC_SLIT_PEAK_NOT_FOUND SIG_CC_SIGNAL_OVERRANGE_REPREFORM_ LAMP_CAL COM_CC_CIRCULAR_BUFFER_OVERRUN _OCCURED 9555 LMP_CC_D2_FILAMENT_FAILURE Foggy Optics PS 335 Error Messages Table can be found: PS 335 Service Manual Dealernet
33
Foggy Optics What is Foggy Optics?
34
What is Foggy Optics? Three types of Foggy optics
White powdery coating Rust/brown spot burn Speckled mirror (delamination)
35
What is Foggy Optics? Diagnosing Foggy Optics
36
Diagnosing Foggy Optics
Run a “wavescan” test, When bad Open the optics and inspect.
37
Wavescan test results Intensity for the 254 nm line
Currently is 10 – 50 K counts
38
False positive New D2 vs Aged D2 Intensities
39
What is Foggy Optics? Diagnosing Foggy Optics Treating Foggy optics
40
0603 Project (Foggy Optics)
Proposed Fixes and their status: White powder fogging: due to leaky flow cells and customer sample depositing on the mirrors New Flow cells & fittings available for all cells except 4 x 0.15 Brown powder fogging: due to UV attack on some of the foam parts close to the UV source. Eliminating as much of the foam as possible and shielding the rest. Mirror delamination: due to UV attack on the thin Al coating of the mirrors Eliminated by putting a NiCr coating on the mirrors to strengthen the Al coating.
41
Opening and cleaning the optics
42
Cleaning 325 optics Clean area in circle Do not adjust!!
43
Wavelength calibration data must be entered into the 335 EEPROM
Replacing Optics Wavelength calibration data must be entered into the 335 EEPROM
44
Replacing Optics This label is attached to the top of each optics module, the shipping box and the test document
45
Summary Single and Dual Array Optical path is open
Flow cell is dual path Diagnostics Tests, like wavescan Shipping Test Main Problems (Foggy Optics)