An evaluation of the isotope ratio performance of an axial TOF - ICP - mass spectrometer F. Vanhaecke, L. Moens, R. Dams, L. Allen* and S. Georgitis* University of Ghent, Laboratory of Analytical Chemistry, Proeftuinstraat 86, B-9000 Gent, Belgium. * LECO Corporation, 3000 Lakeview Ave., St. Joseph, MI 49085 USA.
2
1. Instrumentation LECO Renaissance TOF-ICPMS instrument 123 Acceleration Extraction Skimmer Sampler Detector Ion Mirror ICP Torch Vacuum Stages Flight Tube
3
simultaneous extraction of ions from ICP automatic correction for instabilities in ion source & during ion sampling process on-axis acceleration up to 30000 full spectra / s modified ETP discrete dynode electron multiplier use of analog detection mode to avoid pulse counting saturation no correction for detector dead time required Meinhard type C concentric nebulizer & double pass cyclonic spray chamber (PTFE) free aspiration of sample solution 1. Instrumentation (2)
4
2. Data processing Accurate background correction application of 1 mV offset value
5
use of signal intensity at m/z = 210 as internal reference (no ionic species present) correction for signal drift & instrument instability improvement of both precision & accuracy Isotope ratio precision expressed as RSD% for n = 10 2. Data processing (2)
6
3.Influence of acquisition time per replicate & analyte concentration on precision
7
3.Influence of acquisition time per replicate & analyte concentration on precision(2)
8
Despite use of analog mode, trends in agreement with Poisson counting statistics average improvement 50 500 g/L = 3.5 10 = 3.2 (A) average improvement 1 30 s = 5.2 30 = 5.5 (A) average improvement 10 30 s = 1.8 3 = 1.7 (B) No further improvement on increasing acquisition time > 30 s !! 3.Influence of acquisition time per replicate & analyte concentration on precision (3)
9
4.Typical isotope ratio precision at 30 s acquisition time / replicate (RSD% for n = 10)
10
5.Isotope ratio precision for transient signals (RSD% for n = 5) - FI signals, FWHM = 6 s Isotope ratio precision at 500 µg/L still comparable to the best values ever reported for quadrupole-based ICPMS
11
6. Mass discrimination 1% per mass unit at mid-mass (Cd) comparable to quadrupole-based or sector field instrumentation 7. Accuracy The accuracy attainable was evaluated by comparing Pb isotopic results for a Merck Titrisol ('natural') Pb standard solution obtained by (1) TOF- ICPMS and (2) single collector TIMS. For TOF-ICPMS, 'NIST 981 - Common lead isotopic standard' was used as an external standard.
12
For TOF-ICPMS, the result is the average of 3 subsequent determinations,. Each determination consisted of 5 replicate measurements (30 s) of the 'sample' solution and 5 replicate measurements of the NIST standard solution. For TIMS, the result is the average of 5 determinations. For both techniques, the uncertainties indicated are 95% confidence intervals. 7. Accuracy (2)
13
At sufficiently high signal intensities and for a moderate acquisition time, axial TOF-ICPMS yields a typical isotope ratio precision 0.05% RSD. This precision can be obtained for many ratios simultaneously. The mass discrimination is comparable to that observed with quadrupole-based and sector field instruments and when appropriately corrected for, accurate results can be obtained. 8. Conclusions