1 What is the relationship between near infrared spectroscopic regional mixed arterio-venous oxygen saturation and jugular bulb venous saturation during core cooling and bilateral selective antegrade cerebral perfusion? Vamsidhar B Dronavalli, Ashvini Menon, Ishtiaq Rahman, David Green, Timothy Jones, Jorge Mascaro, Robert S Bonser Department of Cardiothoracic Surgery, University Hospital Birmingham NHS Foundation Trust, Queen Elizabeth Hospital, Birmingham, UK. Conflict of interest - none

2 Aim To assess near infrared spectroscopy (NIRS) as a regional index of mixed arterio-venous cortical saturation (rSO2 ) in managing patients during core cooling (CC)

3 Introduction Increasing jugular bulb venous oxygen saturation(SjO2) during CC correlates with reducing brain O2 consumption and guides safe commencement of hypothermic circulatory arrest (HCA). NIRS a regional index of rSO2 has been advocated as a surrogate of SjO2.

4 Methodology 9 patients undergoing CC Utilising NIRS, rSO2 was measured –At 35 o C, 30 o C, 25 o C and 15 o C during CC –At 5, 15,30 and 45 mins of SACP –At 15 o C, 20 o C, 25 o C,30 o C and 35 o C during warming Utilising a jugular bulb line venous oxygen saturation, S j O 2 was measured at similar time points Standardised anaesthetic and bypass protocol

5 The patients 15 patients Procedures 7x Ascending aorta + arch + elephant trunk ± Aortic valve replacement 2x Ascending aorta+ arch replacement 5x Ascending aortic aneurysm and arch aneurysm 2x Mega aortic syndrome 1X chronic type A dissection 1x Pseudo aneurysm of distal arch and proximal descending Diagnosis Mean age=55

6 Results Temp o CrSO2SjO2SjO2 Core cooling3557±13%77±25% 3061±1391±11%, 2559±1497±3 2067±999±0.7 1575±10%99±0.5% SACP at 15 o C5min 78±697±3 15 min72±1397±4 30 min71±1596±3 End Re-warming phase 35 87±1564±12% Both rSO2 and SjO2 increased with cooling p<0.01 and p=0.03 repectively

7 rSO2 and S j O 2 vs temperature change

8 Scatter plot of SjO2% vs rSO2%

9 Conclusion NIRS rSO2 assessment during CC underestimates SjO 2 for reasons that remain unclear. Regional SO 2 is paradoxically lower than SjO 2 during CC and our data raises questions of what each index of oxygen extraction is actually measuring.