Pipettes Assemble Procedures
This instruction has been prepared to assist you during the maintenance of your HTL pipette. The following diagrams focus on the assemble procedure of the selected (representative) models of the micro, mid-range and macro-volume pipettes. Please follow this instruction during pipette cleaning or replacement of parts of the lower part of the pipette. All information provided here applies to Discovery Comfort / Discovery+, Labmate Pro / Labmate Soft, Optipette and Discovery PRO singlechannel pipette lines.
2
External parts of the pipette:
Pipette Parts „vocabulary” External parts of the pipette: Pipetting Pushbutton + Knob Internal parts of the pipette lower part: Ejector Pushbutton Piston Assay Volume Adjustment Knob + Locking Ring Piston Spring Shaft Nut Compression Set Tip Ejector Compression Bush Shaft Seal and O-ring
3
External parts of the pipette:
Pipette Parts „vocabulary” Pipetting Pushbutton External parts of the pipette: Adjustment Knob (with volume lock mechanism) Internal parts of the pipette lower part: Piston + Spring Ejector Pushbutton Blocking Bush Shaft Cap Compression Spring Shaft [in models DP2 - DP the set consist of: shaft holder set (upper part) shaft (lower part)] Compression Bush Seal and O-ring Tip Ejector