INSTRUCTIONS FOR AEROCHAMBER MINI* AEROSOL CHAMBER (AC)
MD# 969A * Trade marks and registered trade marks of Trudell Medical International. © TMI All rights reserved. We would like to acknowledge Fanshawe College Respiratory Therapy Department for allowing us to use their facility for photography.
2
For Use in a Ventilator Circuit
3
For Use in a Ventilator Circuit
Remove the Metered Dose Inhaler (MDI) aerosol canister from the MDI boot.
4
For use in a Ventilator Circuit
Insert the MDI canister into the AeroChamber mini* AC canister port.
5
For use in a Ventilator Circuit
Prime the MDI prior to use by shaking the MDI and chamber and then depressing the MDI. Repeat this step a second time.
6
For use in a Ventilator Circuit
Disconnect the Ventilator Circuit from the Endotracheal Tube 15mm connector.
7
For use in a Ventilator Circuit
Attach the AeroChamber mini* AC with the canister to the Endotracheal tube 15 mm connector.
8
For use in a Ventilator Circuit
Attach the Ventilator circuit to the 15mm Outside Diameter at the back of the AeroChamber mini* AC.
9
For use in a Ventilator Circuit
While holding the MDI canister vertically, depress the MDI canister downward into the AeroChamber mini* AC actuator and introduce one puff of medication into the Chamber.
10
For use in a Ventilator Circuit
Ventilate the patient with Breaths.
11
For use in a Ventilator Circuit
NOTE: For more than one dose keep the AeroChamber mini* AC in the circuit, remove the MDI canister from the AeroChamber mini* AC, shake the canister and re-insert into the AeroChamber mini* AC canister actuator. Repeat steps 7 through 8.
12
For use in a Ventilator Circuit
Disconnect the AeroChamber mini* AC from the Endotracheal Tube connector and circuit and connect the ventilator circuit to the Endotracheal tube and commence the normal ventilation.