PREX Mtg, July 2008R. Michaels, JLAB Warm Septum : Status Septum Team: Kees deJager, Paul Brindza, Javier Gomez, John Lerose, Al Gavalya, & me New Design choice: 5 0 New energy for PREX: 1.05 0.05 GeV Major engineering design completed Drafting work in progress. Construction to start in Jan 2009, finish by summer 2009
2
PREX Mtg, July 2008R. Michaels, JLAB Septum Magnet target HRS-L HRS-R collimator
3
PREX Mtg, July 2008R. Michaels, JLAB Parameters of Septum B = 0.458 T L eff = 0.873 m N = 380 turns, I = 500 A, NI = 190,000 Amp-turns V = 108 Volts, P = 54 kW 0.4 T-m A pair of dipole magnets right after the target Extends spectrometers to 5 0
4
PREX Mtg, July 2008R. Michaels, JLAB Collimator Septum Magnet Top view
5
PREX Mtg, July 2008R. Michaels, JLAB View from Target Beam Hole Iron Coils Vacuum
6
PREX Mtg, July 2008R. Michaels, JLAB Zoom View from target Issue: How tight a beam hole ? Beam Presently 0.75 0 Copper Coils HRS – L HRS – R Hard limit 1.3 0 (valve) More coils = higher B field for Makis Checking radiation load (Pavel)
7
PREX Mtg, July 2008R. Michaels, JLAB
8
PREX Mtg, July 2008R. Michaels, JLAB Focal Plane X-Y Distributions
9
PREX Mtg, July 2008R. Michaels, JLAB Change in our Acceptance Model
10
PREX Mtg, July 2008R. Michaels, JLAB At 5 0 the new Optimal FOM is at 1.05 GeV (+/- 0.05) (when accounting for collimating small angle, not shown) 1% @ ~1 GeV
11
PREX Mtg, July 2008R. Michaels, JLAB Collimation : define acceptance, cut some low angle (Paul Souder, et.al.)
12
PREX Mtg, July 2008R. Michaels, JLAB Systematic: Location of Collimator no problem, go back to sleep …
13
PREX Mtg, July 2008R. Michaels, JLAB Reduction in Error from Detector Segmentation (Paul Souder) ( 1 st Order Transport )
14
PREX Mtg, July 2008R. Michaels, JLAB Reduction in Error from Detector Segmentation Neutron Radius of Lead
15
PREX Mtg, July 2008R. Michaels, JLAB Analysis of Real Data to find A T Strategy (Paul Souder) Old Septum
16
PREX Mtg, July 2008R. Michaels, JLAB Analysis of Real Data … Focal Plane Distributions Lead Elastic, 1.1 GeV, 6 0 Old Septum
17
PREX Mtg, July 2008R. Michaels, JLAB Same for R-arm
18
PREX Mtg, July 2008R. Michaels, JLAB Analysis of Real Data to find A T Strategy (Paul Souder) New Detector ?
19
PREX Mtg, July 2008R. Michaels, JLAB Warm Septum : Summary New Design choice: 5 0 New energy for PREX: 1.05 0.05 GeV Major engineering design completed Drafting work in progress. Construction to start in Jan 2009, finish by summer 2009