Download presentation
Presentation is loading. Please wait.
Published byLesley McCormick Modified over 9 years ago
1
1 Correct Sampling
2
What is SAMPLING? Intensity [a.u.] 2 3456 X [µm] 1
3
Aliasing … suppose it is a sine-wave Intensity [a.u.] 2 3456 There are many sine-waves, SAMPLED with the same measurements. Which is the correct one?
4
Aliasing … suppose it is a sine-wave … maybe we can know! Object: Microscope Image: Intensity Spatial Coordinate Intensity Spatial Coordinate
5
Aliasing in Fourier-space Fourier-transform of Image Intensity Aliased Frequencies ½ Sampling Frequency ½ Nyquist Frequencies
6
Pixel sensitivity Intensity [a.u.] 2 3456 X [µm] 1 Convolution of pixel form factor with sampling positions Multiplication in Fourier-space Reduced sensitivity at high spatial frequency
7
Optical Transfer Function |k x,y | [1/m] contrast Cut-off limit 0 1 rectange form-factor specimen sampled
8
Consequences of high sampling Confocal: high Zoom more bleaching? No! if laser is dimmed or scan-speed adjusted bad signal to noise ratio? Yes, but photon positions are only measured more accurately binning still possible high SNR. Readout noise is a problem at high spatial sampling (CCD)
9
9 Optimal Sampling?
10
Regular sampling Reciprocal -Sampling Grid Real-space sampling: Multiplied in real space with band-limited information
11
Widefield Sampling In-Plane sampling distance Axial sampling distance
12
Confocal Sampling In-Plane sampling distance (very small pinhole) else use widefield equation Axial sampling distance
13
Confocal OTFs WF 1 AU 0.3 AU in-plane, in-focus OTF 1.4 NA Objective WF Limit
14
Hexagonal sampling Advantage: ~17% + less ‚almost empty‘ information collected + less readout-noise approximation in confocal; 3D: ABA, ABC stacking Reciprocal -Sampling Grid Real-space sampling: Multiplied in real space with band-limited information
15
63× 1.4 NA Oil Objective (n=1.516), excitation at 488 nm, emission at 520 nm eff = 251.75 nm, = 67.44 deg widefield in-plane: d xy < 92.8 nm maximal CCD pixelsize: 63×92.8 = 5.85 µm confocal in-plane:d xy < 54.9 nm widefield axial: d z < 278.2 nm confocal axial: d z < 134.6 nm Fluorescence Sampling Example
16
OTF is not zero but very small (e.g. confocal in-plane frequency) OTF is not zero but very small (e.g. confocal in-plane frequency) Object possesses no higher frequencies You are only interested in certain frequencies (e.g. in counting cells, serious under-sampling is acceptable) Reasons for Undersampling
17
Detector generates high-frequency noise? Detector generates high-frequency noise? Measure this noise (e.g. dark exposure and 2D FFT) Avoid aliasing by sampling above this noise frequency. Traps and Pitfalls
18
FFT of dark CCD exposure (2 µs)
19
If you need If you need high resolution or need to detect small samples sample your image correctly along all dimensions sample your image correctly along all dimensions Sampling Summary
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.