PAT TESTER CALIBRATION Using the 3200 Electrical Test Calibrator.

Similar presentations


Presentation on theme: "PAT TESTER CALIBRATION Using the 3200 Electrical Test Calibrator."— Presentation transcript:

1

2 PAT TESTER CALIBRATION Using the 3200 Electrical Test Calibrator

3 Dedicated inputs and test buttons for PAT Testers provides safe, convenient connection & operation The 3200 has six functions which can be used for testing of Portable Appliance Testers (PATs) : 1. Earth Bond Resistance measurement 2. Earth Bond Current measurements 3. Insulation testing 4. Load testing 5. Flash Testing 6. Leakage Testing PAT TESTER CALIBRATION Overview

4 1.CONNECT TESTER TO PAT EARTH BOND TERMINAL AND PAT TEST IEC SOCKET 2.SET EARTH BOND RESISTANCE FUNCTION ON TESTER 3.SET EARTH BOND RESISTANCE RANGE ON 3200, EG. 1 OHM 4.START EARTH BOND MEASUREMENT ON PAT TESTER PAT TESTER CALIBRATION PAT Earth Bond Resistance

5 PAT TESTER CALIBRATION PAT Earth Bond Current 1.CONNECT PAT TO 3200 EARTH BOND TERMINAL USING SPECIFIC EARTH BOND CURRENT TERMINAL & CONNECT TO PAT TEST IEC SOCKET 2.SET EARTH BOND FUNCTION ON TESTER 3.SET EARTH BOND CURRENT RANGE ON 3200, EG. 10A AND CLICK ‘TEST’ SOFTKEY 4.START EARTH BOND MEASUREMENT ON PAT TESTER – WAIT FOR 5 SECOND TEST INTERVAL

6 PAT TESTER CALIBRATION PAT Insulation Resistance 1.CONNECT PAT TO 3200 EARTH BOND TERMINAL & CONNECT TO PAT TEST IEC SOCKET 2.SET INSULATION RESISTANCE MEASUREMENT FUNCTION ON TESTER 3.SET PAT INSULATION RANGE ON 3200, EG. 9M OHMS 4.START INSULATION RESISTANCE MEASUREMENT ON PAT TESTER

7 PAT TESTER CALIBRATION PAT Load Testing 1.CONNECT PAT TO 3200 EARTH BOND TERMINAL & CONNECT TO PAT TEST IEC SOCKET 2.SET LOAD / RUN TEST FUNCTION ON TESTER 3.SET PAT LOAD MODE AS REQUIRED (O/CS/C0.13kVA) 4.START MEASUREMENT ON PAT TESTER

8 PAT TESTER CALIBRATION PAT Leakage Current 1.CONNECT PAT TO 3200 EARTH BOND TERMINAL & CONNECT TO PAT TEST IEC SOCKET 2.SET LEAKAGE CURRENT MEASUREMENT FUNCTION ON TESTER 3.SET LEAKAGE RANGE ON 3200, EG. 2mA 4.START LEAKAGE MEASUREMENT ON PAT TESTER SUBSTITUTE LEAKAGE MEASUREMENT CAN ALSO BE PERFORMED AT 40V TEST LEVELS USING THE 3200