Presentation is loading. Please wait.

Presentation is loading. Please wait.

Systematics on ScECAL 1 st prototype DESY Apr 18-21 2015, CALICE Satoru Uozumi for the CALICE collaboration NIM A, 763 (2014) 278.

Similar presentations


Presentation on theme: "Systematics on ScECAL 1 st prototype DESY Apr 18-21 2015, CALICE Satoru Uozumi for the CALICE collaboration NIM A, 763 (2014) 278."— Presentation transcript:

1 Systematics on ScECAL 1 st prototype test @ DESY Apr 18-21 2015, CALICE meeting@KEK Satoru Uozumi for the CALICE collaboration NIM A, 763 (2014) 278

2 MPPCs (1600 pixels) The ScECAL 1 st prototype Scintillator strip (1 x 4.5 x 0.3 cm) Acryl Frame WLS fiber Tungsten (3.5 mm thick) Scintillator layer (3 mm thick)

3 Fibre and Direct readout Modules ① WLSF readout 13 layers ② Direct readout 13 layers Beam DFFD 13 + 13 layers F-type D-type Longitudinal response uniformity for MIP F-type D-type

4 1-6 GeV e + beams into Central & Uniform regions Detector front view At DESY, 1 st prototype had Been tested on 2007 using 1-6 GeV e + beam. Detector surface are categorized into Central region less EM shower leak, more effect of response non-uniformity Uniform region (all violet squares) little shower leak less effect of non-uniformity

5 Linearity & Energy resolution

6 Resolution on Systematics : Breakdown Derived from the e + beam : Not easy to estimate The beam momentum spread precisely Reasonably small Improved in 2 nd prototype Non-trivial

7 Systematics : Optical cross-talk

8

9 The Mega-strip structure had been Replaced with separated strips after 2 nd prototype.

10 Systematics : Single Pixel signal

11 AHCAL electronics has 2 gain modes. LED calibration has been done in 2 steps in order to obtain the single pixel equivalent signal. d-value Step 1 : Measure single-photoelectron peak in high gain mode, Step 2 : Measure gain ratio of low/high-gain mode (=inter-calibration factor). Note : LED runs are done just for a limited number of channels

12 Systematics : Single Pixel signal Due to a lack of available time, Inter-calibration had been done just for small fraction of whole channels. This incompleteness of LED runs induces non-negligible systematic uncertainties. It tells the necessity of well-arranged LED system & runs. Inter-calib factor for 30 channels (out of 468). ~10% of uncertainty

13 Systematics : Effective Pixel Number

14 Systematics : effective pixel number Difference of those Is taken as the source of systematic uncertainty Non-uniform response of the MPPC is corrected using correction curve (measured separately) : Thanks to quick recovery (~4ns) of the MPPC, num. of effective pixels is enhanced. The enhancement depends on input light length. This tells that larger dynamic range and knowledge on response of the MPPC is important.

15 Summary & Conclusion From the viewpoint of irreducible systematics, Beam momentum spread gives dominant effect Understanding non-linear response of the MPPC is important Single-pixel & electronics calibrations are of 2 nd importance (balance with technology & cost) Strip response non-uniformity also gives non-trivial effect MIP calibration, temperature dependence of the MPPCs are rather OK

16 Backups

17 Feedbacks to 2 nd & future prototypes The 2 nd ScECAL prototype had been built and tested @ FNAL (1-32 GeV e-, p- and m)

18 The DESY Beam-line ECAL module Movable stage e + 1~6 GeV Veto1 Drift chambers DESY II Electron Synchrotron e-e- Drift chambers Trigger-1 Trigger-2 & Veto-1 ECAL Veto-2 e + beam T1 T2 Veto2 T3

19

20


Download ppt "Systematics on ScECAL 1 st prototype DESY Apr 18-21 2015, CALICE Satoru Uozumi for the CALICE collaboration NIM A, 763 (2014) 278."

Similar presentations


Ads by Google