Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul, Korea 2 2 Sae Yan Eye Clinic, Seoul, Korea 3 3 Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea. None of the authors have any commercial or financial interests
2
For Successful Outcome Of Toric IOL Precise Measurement Of Preoperative Corneal Astigmatism Crucial Point ! To Determine Accuracy & Correlation of 6 Different Keratometers & Topographers 2
3
Shin-Nippon Co, Ltd, Japan Measuring capacity Corneal Radius of Curvature ○ 6.50 ~ 9.40 mm Corneal Keratometric Power ○ 36 ~ 52 D Axis Corneal Astigmatism ○ 0 ~ 180° 3
4
Canon, Japan Measuring capacity Corneal Radius of Curvature ○ 5.5 ~ 10 mm Corneal Refractive Power ○ 33.75 ~ 61.25 D Axis Corneal Astigmatism ○ 0 ~ 180° 4
5
Carl Zeiss Meditec, Germany 2.5 mm measurement diameter Cf. Manual keratometers ~ 3 mm 30 data points per measurement 6 spots x 5 samples 5 images automatically captured in 0.5 sec Displayed result is average of 5 captured images 5
6
6 Bausch & Lomb, USA Topography-based Simulated K-readings (=Sim K’s) c Axis 3 & 5 mm zone statistics Average & SD of Mean & Astigmatic Power Steep & Flat Axis
7
Oculus, Germany Scheimpflug image-based Topographic analysis of Corneal Astigmatism 7
8
Tracey Technologies, USA Dual Acquisition Unit EysSys Corneal Topographer ○ Placido ring-based Wavefront analyzer 8
9
Study Population Preoperative visit for Cataract Surgery 257 eyes of 141 patients ○ Age : 54.5 ± 20.5 yrs ○ Sex : Male - 54, Female - 87 Manual Keratometric Measurement Duplicated examination by 2 skillful examiners Considered as Standard Compared with 5 other Keratometers 9
10
Pearson Correlation analysis Between Manual Keratometery Vs Other Methods 10 Vs. Manual Keratometer Pearson Correlation Coefficient iTrace0.947 ORBscan0.921 AutoKeratometer0.919 IOL master0.857 Pentacam0.847
11
Pearson Correlation analysis Between Manual Keratometery Vs Other Methods 11 Vs. Manual Keratometer Pearson Correlation Coefficient iTrace0.856 AutoKeratometer0.852 IOL master0.751 ORBscan0.646 Pentacam0.633
12
Duplicated Manual Keratometric Measurements By 2 Experienced Operators Taken As Standard iTrace Approximated This Standard The Most Auto Keratometer, ORB & IOL Master Showed To Be Reasonably Accurate. Trend For ORB Fail To Read Accurately In Steep Axis 12