Casper Maheshwari advisors: Stephen Ulhorn & E. Duco Jansen Department of Biomedical Engineering Vanderbilt University http://vubme.vuse.vanderbilt.edu/casper/home.htm Design of an OCT Probe
2
Project Description zDesign, build, and implement a palm-size probe that can be used at the sample arm of an OCT imaging system. zThis probe will be used to differentiate between normal and tumor brain tissue.
3
What is OCT? zOptical Coherence Tomography is an imaging modality that produces 3-D images of biological microstructures. zThe heart of the OCT system consists of a Michaelson Interferometer. ySample Arm (lateral scanning) yReference Arm (longitudinal scanning) zThe laser travels through the sample and reflects off biological microstructures. y(depth scanning)
4
OCT Block Diagram LUMINESCENT DIODE sample 50/50 BANDPASS FILTER ENVELOPE DETECTOR A/D COMPUTER LATERAL SCANNING LONGITUDINAL SCANNING PROBE MODULE
5
ParametersParameters zDefined Wavelength = 1.3 m yTotal Optical Path Distance = 49.0 cm yDepth of Focus between 3.0-5.0 mm yLateral Scanning Distance = 5.0 mm zCalculated yCollimating Lens: f = 4.4 mm xBeam diameter = 1.15 mm yFocusing Lens: f = 40 mm xDepth of Focus = 4.0 mm yFiber Patch Cord Length = 29.31 cm
6
Proposed Probe Design scanner Fiber l = 29 cm collimating lens f = 4.4 mm focusing lens f = 40 mm sample 5 mm 40 mm 1.15 mm diameter
7
Completed Work z Library research of already existing probes z Arrival of the OCT system parts z Assembly of the OCT system z Basic probe design z Calculation of probe parameters z Order of probe components
8
Current Work z Awaiting Arrival of Probe components ycollimating and focusing lenses ypatch fiber cord Ray Tracing on Optics Lab ®
9
Costs z Fiber Patch Cord$ 51.00 z Collimating Lens$ 44.00 z Focusing Lens$ 88.00 z Total (to date)$ 183.00
10
Future Work & Problems z Future Work yconstruct a probe prototype ydesign and make housing for the probe ytest out probe with basic experiments z Problems yThe desired scanner takes 10-12 wks to arrive yAspheric collimating lens unavailable