CMS HF PMT SYSTEM By Y. ONEL U. of Iowa, Iowa City, IA CMS HCAL at Fermilab Feb 6-8, 2003
2
CMS-HF PMT Test and Quality Control System U. Akgun 1, A.S. Ayan 1, F. Duru 1, E. Gulmez 2, M. Miller 1, J. Olson 1 Y. Onel 1, I. Schmidt 1 with Quarknet Group – P. Bruecken, C. Like, R. Newland 1 University of Iowa, Iowa City, USA 2 Bogazici University, Istanbul, Turkey Abstract We have measured the specifications proposed by the CMS-HCAL committee on the candidate phototubes from the three major manufacturers; Hamamatsu, EMI and Photonis. In this report, we present the results from those measurements and we outline the future measurements for the test and the quality control as well as the design of the new University of Iowa PMT test station facility.
3
Tasks of the Test System For one tube in every batch: Double-pulse linearity, Gain vs HV for each batch Single photoelectron spectrum X-Y scan (spatial uniformity) Lifetime For each tube: Pulse width Pulse rise time Transit time Transit time spread Anode dark current Relative gain coupled with cathode sensitivity, Pulse linearity Quality control decision on each tube.
4
Iowa PMT Timing Test Setup
5
UNIVERSITY of IOWA PMT TEST STATION
6
XY Uniformity, Dark Current, Relative Gain 32 channel Voltage ADC Motor Controller LabVIEW Software SCSI Bus Serial Bus 8-PMT Array Laser Pinhole Mask Camac Sys 8 channel pico-ammeter
7
Single Photoelectron Setup (fourth generation)
8
LabVIEW software
9
PMT Timing Data (1900 PMT’s)
11
PMT Data (1900 PMT’s)
12
CA0058 Double Pulse Linearity
13
Single Photoelectron Spectrum at 1100V
14
Single Photoelectron Spectrum at 1500V
15
XY Uniformity
16
Relative Gain vs Gain
17
Gain vs HV for Relative Gain %50-%70
18
Gain vs HV for Relative Gain %70-%80
19
Gain vs HV for Relative Gain %80-%90
20
Relative Gain vs Relative QE
21
Relative Gain & Relative QE
22
Lifetime Measurement Setup
23
Timing characteristics after 1100 C 0472 Pulse WidthRise TimeAv. Transit TimeTransit Time Spread Before 3.74ns2.02ns15.5ns0.148ns After 3.74ns2.14ns15.4ns0.173ns 0252 Pulse WidthRise TimeAv. Transit TimeTransit Time Spread Before 4.12ns1.98ns15.5ns0.094ns After 3.8ns2.12ns15.4ns0.174ns
24
PMT Web Database Sort by column (Ascending or Descending) Pagination reference for large data sets Alternating colors to aid readability More extensive search/sort options are being developed
25
PMT Web Database
26
PMT Base Comparison Hamamatsu’s Resistive Base Cockroft-Walton Base Parallel-Dynode Bases
27
Preliminary Base Comparison Results
29
HCAL LED Pulser
30
HF LED Pulser
31
HF LED Pulser Output of Prototype 10 ns/div Electrical Signal Degradation at End of 10 cm cable Original Pulse at LED
32
Preliminary SPES of LED
33
HF LED Pulser (next generation) - Reduced pulse width - Decreased rise time
34
HF PMT Papers
35
CMS Notes CMS IN 2002/026 CMS IN 2002/032
36
CMS Notes CMS IN 2002/030 CMS IN 2002/029
37
Manufacturer specIowa Tests Window Material Borosilicate glass PASSNA Eff. Pho.cath. dia. 22-28mm, head-on PASSNA Quantum efficiency >15% 400-500 nm PASSNA Photocathode lifetime >200 mC PASSNA Anode current vs position <+/-20% with 3 mm spot scan PASS Gain 10^4 to 10^5,10^5 at <0.75 x V ka(max) PASS Single pe resolution rms/mean if single pe peak 50% or better PASS Pulse linearity +/- 2% for 1-3000 photoelectrons (g=4X10^4) PASS Anode pulse rise-time <5ns PASS Transit time <25 ns preferred PASS Transit time spread <2 ns preferred PASS Anode pulse width <15 ns FWHM PASS Gain (1/2)-lifetime >1500 C PASSNA Gain recov. (2000pe pulse) within 10% of nominal (g=10^4) in 25 ns PASS Average current Ik <1 nA (g=10^4) PASS Average current Ia <10 microA (g=10^4) PASS Anode dark current <2 nA (g=10^4) PASS Stability <+/- 3% within any 48 hr. period PASSNA Envelope opaque and -HV conductive coating PASSNA