Current Status of LASER FRAME for KEK-Nano BPM (Tentative results of resolution test) Second Mini-Workshop on Nano Project at ATF December 11-12, 2004 KEK-Nano BPM Group Y.Higashi, Y.Honda, T.Tauchi, H.Hayano, J.Urakawa, T.Kume,K.Kubo, H.Yamaoka…..
2
Outline Why we need Laser Frame Complete design of the Laser Frame Configuration Laser BPM Assembly Tentative results of resolution test Summary
3
Why we need Laser Frame Initial alignment (10 m) OD of cavities will be used as reference Fine alignment using BPMs signals Needs Nano-Movers Needs stable positioning with nm order Needs reference lines Laser Frame
4
Concept of Laser Frame We use; @ Laser-BPM @ Interferometer @ Vacuum environment * Eliminate the residual errors in Laser interferometry from air turbulence and nonlinearity
5
Over view of KEK Nano BPM
6
Vertical Interferometer Extended Reference Bar Reference Bar Plane Mirror Leg Vacuum pipe, chamberB.S, Mirror, Detector Laser BPM Two Beam Optics Laser Beam from Fiber
7
Laser BPM Reference Bar L 2L dy P1 dP1=2Lsin +dy dP2=Lsin +dy sin dP1-dp2)/L dy= dP1+Lsin D11 D12 D21 D22 P1=D11-D12 P2=D21-D22
8
Cross sectional drawing of a Laser BPM (including vacuum chamber) To detector Vacuum chamber Laser ray Ground
9
Resolution test setup conditions => Laser: YAG Laser (CW,500mW, 532nm, Single mode, W 0 =0.9mm, Divergence 1mrad.) =>Environment (not vacuum but transport in the pipe ) =>BPM location (2 m distance from the reference beam generator) =>Beam splitter ( PBS 50% transmission) =>Detector (diff. amp gain 100-10000) =>Base (Granite table 1x2m. 0.3m thick no apply vibration isolator) =>BPM movement ( measured by capacitance gauge(resolution: 3nm) =>Tilt measurement ( light lever racio 1:2)
10
Test items Use a Single Laser BPM =>Vertical resolution =>Tilt resolution
11
Calibration setup of beam position measurment(1) 2 m Laser 4 th BPM Reference beam generator
12
YAG Laser 500mW Optical fiber Beam divider 1 st BPM 2 nd BPM 3 rd BPM Setup (2) Beam intensity of n th BPM= p * 0.5 n (n= BPM number, p=power)
13
Vacuum chamber Laser ray Resolution test of the 4 th BPM
14
Resolution test Results Laser power: 500 mW, 532 nm micrometer stage 1 um ~ 1 V (Amp. gain 2000) ⇒ 1nm~1mV Spot size = 2900 um (1-sigma,at 2m from source point) w 0 = 900 um (divergence 1mrad.) Rayleigh length = πw 0 ^2 /λ= 4.8 m
16
Summary Vertical resolution => 10 nm Tilt resolution => 5x10 -6 rad. Need to Test => Long distance (1-2 m) Interferometer
17
Elevation View
18
Reference Bar
19
Laser Input
20
Distance change between Ref. Bar and cavity (Due to the environmental change) 7 m 1 hour
21
Deviation due to assembling and fabrication errors (2Kgf) dY +0.1 m-0.4 m -0.6 m+0.4 m -0.1 m