HBD HV Control and Monitoring System RESISTORS, CALIBRATION & CO. Manuel Proissl HBD Meeting 12/23/2008
2
OUTLINE Project: dV = 0 Step 1: Calibration of LeCroy PS Step 2: Resistors on Divider Step 3: Mesh Voltage Sweep vs. ADC Readout Project Progress Calibration Routine / Curves Calibration Program in HV Control Divider(s) Resistor Measurements on HBD Cable Coding (…) Coming Up Upgrades For Run9 | PHENIX, BNL HBD HV Control System HBD Meeting :: December 23 rd, 2008 HBD HV Control & Monitoring System Manuel Proissl Page 2 of 6
3
PROJECT: dV = 0 Step 1: In order to accomplish the project’s goal, the fundamental prerequisite is knowing the ACTUAL voltage we’re applying… Issue:Difference between DV, MV (mainframe) & actual voltage supplied Solution:Calibration Curves for all channels Step 2: What voltage ends up being supplied to the GEM/Mesh? Issue:Resistor values only known by color code (…) Solution:Measurement of all resistors installed on divider board Step 3: What is going to be The Mesh Voltage for RB? Issue:Who knows? Solution:1+1 Mesh Voltage Sweep during +/- config and ADC readout files for every sweep voltage analysis peak of PE vs. Mesh V. HBD HV Control & Monitoring System Manuel Proissl Page 3 of 6 Upgrades For Run9 | PHENIX, BNL HBD HV Control System HBD Meeting :: December 23 rd, 2008
4
PROJECT PROGRESS Under IR Access > MF Calibration, HV Control runs a calibration procedure automatically for every channel, while the Fluke via divider records every second (FlukeView) the output voltage of the particular channel on the LeCroy After initial non-linearity problems (unknown current development), further calibration work will be done (…) ALL resistors have been measured with Fluke in nS through FlukeView with 20 stable events each for average calculation These will be out shortly… available in the hvdb database ALL take-out resistors have been labeled with an unique position code From now on, any future changes can be done safely without any mix-up! HBD HV Control & Monitoring System Manuel Proissl Page 4 of 6 Upgrades For Run9 | PHENIX, BNL HBD HV Control System HBD Meeting :: December 23 rd, 2008
5
THE RESISTOR CODEC HBD HV Control & Monitoring System Manuel Proissl Page 5 of 6 EW View from IR Front ES3 ES4 ES5 EN3EN5 EN4 WS3 WS4 WS5 WN3WN5 WN4 ES1 ES2 EN2 EN1 WS1 WS2 WN2 WN1 SNSN 13 2 57 468 Resistor ID Code HV Box (1-8) Module Num. (1-5) GEM (T, M, B) Layer (1-3) Upgrades For Run9 | PHENIX, BNL HBD HV Control System HBD Meeting :: December 23 rd, 2008
6
COMING UP Resistor Values into hvdb database Connecting to config system in HV Control HV remote testing until Jan 2 nd Correction Curves for 1471N Calibration: new divider setup… Distribution / Installation Files for Linux & XP/Vista Platform independence Final installation files with Installation Wizard HV remote testing / system improvements until Jan 2 nd HAPPY HOLIDAYS!!! HBD HV Control & Monitoring System Manuel Proissl Page 6 of 6 Upgrades For Run9 | PHENIX, BNL HBD HV Control System HBD Meeting :: December 23 rd, 2008