TAbMEP Assessment: POLARCAT H 2 O Measurements. Assessment Summary Table 1. Recommended POLARCAT H 2 O measurement treatment AircraftInstrument Reported.

Similar presentations


Presentation on theme: "TAbMEP Assessment: POLARCAT H 2 O Measurements. Assessment Summary Table 1. Recommended POLARCAT H 2 O measurement treatment AircraftInstrument Reported."— Presentation transcript:

1 TAbMEP Assessment: POLARCAT H 2 O Measurements

2 Assessment Summary Table 1. Recommended POLARCAT H 2 O measurement treatment AircraftInstrument Reported 2σ Uncertainty Recommended Bias Correction Recommended 2σ Uncertainty NASA DC-8DLH5%-0.0562 - 0.00082 H 2 O DC8 {(-0.0562 - 0.00082 H 2 O) 2 + (.07) 2 } 1/2 NOAA WP-3DAOC5%- 0.0491 - 0.0950 H 2 O WP3D {(- 0.0491 - 0.0950 H 2 O) 2 + (0.308) 2 } 1/2 NASA P-3B 10%- 0.0903 + 0.0275 H 2 O P3B {(- 0.0903 + 0.0275 H 2 O) 2 + (.834) 2 } 1/2 DLR FALCON 10%- 0.0724 + 0.0673 H 2 O DLR {(- 0.0724 + 0.0673 H 2 O) 2 + (0.166) 2 } 1/2 ATR-42 FALCON 10%0.275 - 0.0130 H 2 O ATR42 10% a a The PI reported uncertainty was recommended because no IEIP precision value could be calculated for the ATR-42 Falcon data

3 Assessment Summary Cont. Intercomparison Data Within the Recommended Uncertainty DC-8 vs. WP-3D45% DC-8 vs. P-3D62% DC-8 vs. DLR94% DLR vs. ATR-4276% Only the DC-8 vs. DLR Falcon intercomparison meets the condition that 90% of data fall within the 2σ uncertainty bounds, making it a TAbMEP unified data set Remaining intercomparisons contain systematic error that cannot be fully accounted for in uncertainty calculations, causing the recommended uncertainties to be insufficient for TAbMEP unified data set requirements Table 2. TAbMEP unified data sets

4 Assessment Summary Cont. Table 3. POLARCAT H 2 O bias estimates AircraftInstrument Apparent Bias (a g/kg + b H 2 O) Best Estimate Bias (a g/kg + b H 2 O) NASA DC-8DLH0-0.0562 - 0.00082 H 2 O DC8 NOAA WP-3DAOC0.00649 - 0.0861 H 2 O DC8 - 0.0491 - 0.0950 H 2 O WP3D NASA P-3B - 0.0351+ 0.0291 H 2 O DC8 - 0.0903 + 0.0275 H 2 O P3B DLR FALCON - 0.0174 + 0.0731 H 2 O DC8 - 0.0724 + 0.0673 H 2 O DLR ATR-42 FALCON 0.327 - 0.0118 H 2 O DC8 0.275 - 0.0130 H 2 O ATR42

5 Assessment Summary Cont. FlightPlatform IEIP Precision Expected VariabilityObserved Variability Adjusted Precision 4/12 DC-82.2% 15.6% 2.2% WP-3D 15.4% 4/8 DC-81.5% 12.5% 1.5% P-3B 12.4% 4/19 DC-83.5% 41.8% 3.5% P-3B 41.7% 7/10 DC-82.5% 7.85% 2.5% P-3B 7.4% 7/9 DC-82.5% 8.66% 2.5% DLR Falcon 8.3% 4/15 P-3B 9.63% WP-3D 7/14 DLR Falcon 20.5% ATR-42 Falcon Table 4. POLARCAT H 2 O precision (1σ) comparisons

6 P-3B vs. DC-8

7 P-3B vs. DC-8 04/08/2008

8 DC-8 04/08/2008 IEIP = 1.5%

9 DC-8 04/08/2008 STD = 0.0022 g/kg

10 P-3B vs. DC-8 04/08/2008

11

12

13 P-3B vs. DC-8 04/19/2008

14 DC-8 04/19/2008 IEIP = 3.5%

15 DC-8 04/19/2008 STD = 0.0045 g/kg

16 P-3B vs. DC-8 04/19/2008

17

18

19 P-3B vs. DC-8 07/10/2008

20

21 DC-8 07/10/2008 IEIP = 2.5%

22 DC-8 07/10/2008 STD = 0.0050 g/kg

23 P-3B vs. DC-8 07/10/2008

24

25

26 P-3B vs. DC-8 All Flights

27

28

29 WP-3D vs. DC-8 04/12/2008

30

31 DC-8 04/12/2008 IEIP = 2.2%

32 DC-8 04/12/2008 STD = 0.0030 g/kg

33 WP-3D vs. DC-8 04/12/2008

34

35

36 P-3B vs. WP-3D 04/15/2008

37

38

39

40

41 DLR Falcon vs. DC-8 07/09/2008

42

43 DC-8 07/09/2008 IEIP = 2.5%

44 DC-8 07/09/2008 STD = 0.0050 g/kg

45 DLR Falcon vs. DC-8 07/09/2008

46

47

48 DLR Falcon vs. ATR-42 Falcon 07/14/2008

49

50

51

52

53