High Level Disinfection of Clinical Equipment and Strategies for Endoscopes
Larissa Lewis, RN, BSN, CIC Infection Preventionist UW Harborview Medical Center
2
Why is it important?
4
Reported Outbreaks from Duodenoscopes
Florida 2008 Pittsburg 2012 Chicago 2013 Los Angeles Seattle
5
Who’s Interested? …And anyone else who might accredit your facility
6
Who Offers Guidance They don’t always agree!
7
The basics
8
A Perfect World Single Use or Single Patient Use
Reasons why this is not possible: Expense Waste Practicality
9
Spaulding Classification
Earl Spaulding was a microbiologist at Temple University Over 30 years ago he proposed classification of patient care items and medical equipment and appropriate level of disinfection Earl Spauling wrote a paper on disinfection in 1939 using these:
10
Non-Critical Equipment/ instruments that contact intact skin
11
Semi-Critical Equipment/ instruments that contact mucous membranes or nonintact skin
12
Critical Equipment/ instruments that enter sterile tissue or the vascular system
13
Clean Removal of soil
14
Disinfection elimination of most pathogenic microorganisms(not spores)
15
Disinfection Low-level disinfection – High-level disinfection –
elimination of many pathogenic microorganisms High-level disinfection – elimination of all pathogenic microorganisms except spores
16
Sterilization elimination of all pathogenic microorganisms, including spores
17
Back to Earl Spaulding…
Critical Items Equipment/ instruments that enter sterile tissue or the vascular system Surgical equipment Implants Sterilization Semi-critical Items Equipment/ instruments that contact mucous membranes or nonintact skin Many endoscopes Vaginal US probes Respiratory therapy equipment High level disinfection Noncritical Items Equipment/ instruments that contact intact skin Blood pressure cuff Patient environment Ambulation aids Low level disinfection
18
DON’T FORGET TO CLEAN! Cleaning is the fundament to disinfection regardless of the level!
19
Why? Biofilm! FEMS Microbiol Rev 36 (2012) 972–989
20
Cleaning ultrasound probes vs scopes
These are both semi-critical items Vs.
21
Sterilization Steam Hydrogen peroxide gas plasma
100% Ethylene oxide (ETO) ETO mixtures Vaporized hydrogen peroxide Ozone
22
High Level Disinfection
Peracetic acid/hydrogen peroxide Gluteraldehyde Hydrogen peroxide Ortho-phthalaldehyde (OPA) Peracetic acid
23
Manual High Level Disinfection
Active times vary by product; 7-90 minutes Chemicals are inactive below certain temps, vary by product; 20-30°C Expiration dates of chemicals Overall expiration Expiration of the product once open Expiration of the product in use May also have to test solution for quality and keep track of test strip Lots of subtly and variation between products
24
Lots to Track!
25
Automated High Level Disinfection
Completes all steps of high level disinfection following manual cleaning Often device specific Most often used with endoscopes Device compatibility is essential
26
High Level Disinfection is EVERYWHERE!
Outpatient clinics/procedure areas Inpatients ancillary departments Inpatient procedure areas Anywhere with Scopes Ultrasound probes Respiratory equipment/ anesthesia equipment
27
Why is it so important to know where high level disinfection is being done?
28
This! All staff must be trained and competency validated
All staff must be trained and competency assured.
29
And this…
30
Additional Monitoring for Endoscope Procedures
ATP? – adenosine triphosphate monitoring Measures “cleanliness” Culturing? – take cultures off scopes after procedures Surveillance? – track patients who’ve had the procedure and monitor for infection
31
Most important Know where you have high level disinfection
Ensure device/chemical compatibility Ensure device/automated reprocessor compatibility Make sure your peeps are trained Track EVERYTHING Follow manufacturer’s directions The scope The chemicals/automated reprocessor compatibility