1
Sysmex XE-2100 Counting of Red Cells and Platelets by impedance & fluorescent techniques
2
Methods employed Red cells are measured by impedence only Platelets are measured in two channels – Impedance Channel (default) – Fluorescent Channel Impedance Channel is default count In instances of abnormality when the algorithm will come into play
3
Impedance Channel - Overview Red cells and platelets Hydrodynamic Focusing of cells to be counted DC aperture Impedance Excellent Linearity – with limitations
4
Impedance Counting Platelet count is fine when histograms are well defined But in case of abnormality, platelet count suffers from Rbc interference PLT_DC RBC_DC
5
Optical PLT Count In those instances - Fluorescent Optical Platelet count would be preferable for several reasons RET(Plt-O)
6
Optical System Higher sensitivity due staining of cells Better Rbc/plt resolution Necessary in – Small (Microcytic) RBC – Giant Platelets – Rbc Fragments
7
Measurements used in Optical System Forward Scatter - size of cells Side Scatter - internal structure (i.e size of nucleus) Fluorescence - Measurement of RNA/DNA stained Fluorescent Optical Platelet count is performed with reticulocyte count
8
In the presence of Rbc fragments, microcytic RBC or giant platelets, the PLT distribution curve is clearly affected However in the RET scattergram, the RBC and PLT-O clusters are obviously distinct Accurate Platelet Counting
9
Limitations of Optical Platelet Counts WBC fragments are included in the Optical Platelet Count Generates a “Plt Abn Scattergram” flag
10
PLT-IPLT-O WBC fragments / giant platelets: PLT-IPLT-O RBC fragments/ small Rbcs: PLT Switching Algorithm
11
PLT-I, Low reliability PLT-O, Low reliability PLT Vote out PLT-I is reported PLT-O is reported No Yes PLT Switching Algorithm
12
PLT Abn Distribution flag or Fragments? flag is detected from RBC channel or PLT-I/PLT-O count differences +/- 15% Changing the count?
13
The PLT-I/O Switching Rule For low-PLT# – PLT-O# may more reliable than PLT-I#, when PLT-I# are low Customizable switching trigger for PLT-I# – Default setting is 50.0x10^3/uL
14
Adding PLT-I/O switching rule