1
Example of Aliasing
2
Sampling and Aliasing in Digital Images Array of detector elements Sampling (pixel) pitch Detector aperture width The spacing between samples determines the highest frequency that can be imaged Nyquist frequency: F N = 1/2 If a frequency component in an image > F N → sampled F N → sampled < 2x/cycle: aliasing Wraps back into the image as a lower frequency Moiré pattern, spoke wheels c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 284.
3
Sampling and Aliasing in Digital Images Example: sampling pitch of 100 m → F N = 5 cycles/mm When input f > F N then the spatial frequency domain signal at f is aliased down to: f a = 2F N – f Not noticeable with patient Antiscatter grids c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., pp. 285-286. Aperture blurring - signal averaging across the detector aperture
4
Aliasing due to Reciprocating Grid Failure
5
Noise is anything in the image that is not the signal we are interested in seeing. Noise can be structured or Random.
6
Structure Noise Noise which comes from some non-random source: breast parenchyma, hum bars in CRT’s. The design goal in making an imaging system is to reduce structure or system noise to below the level of the random noise.
7
Random or Quantum Noise Noise resulting from the statistical nature of the signal source is random or quantum noise. In imaging, the signal is light in the form of photons being emitted randomly in time and space. Because we are working with a random source, we can use statistics to describe the behavior of the image noise.
8
Rose Model The information content of a finite amount of light is limited by the finite number of photons, by the random character of their distribution, and by the need to avoid false alarms (false positives). The measure of how well an object (signal) can be seen against a background of varying signal strength (noise) is the signal to noise ratio: S/N.
9
Rose Model To see an object of a given diameter (resolution) you must have sufficient contrast and S/N. In an ideal system, where the only noise is quantum noise, the diameter, D, which can be resolved is given by: D 2 x n 2 = k 2 /C 2 where C is the contrast of the detail, n is the number of photons/sq cm in the image, and k is the threshold S/N ratio. Most people use k=5. (remember, good resolution means D is small)
10
Contrast Resolution Ability to detect a low- contrast object Related to how much noise there is in the image → SNR As SNR ↑ the CR ↑ Rose criterion: SNR > 5 to reliably identify an object Quantum noise and structure noise both affect the conspicuity of a target c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 281.
11
Statistics as image models
12
Gaussian Probability Distribution Function Gaussian (normal) distribution: the mean the mean and σ describe the shape Many commonly encountered measurements of people and things make for this kind of distribution (Gaussian) hence the term “normal” e.g., the height of 1000 third grade school children approximates a Gaussian c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 275.
13
FOR GAUSSIAN PROBABILITY DISTRIBUTION MEAN X = X i i N VARIANCE 2 = i ( X i - X ) 2 (N - 1)
14
FOR GAUSSIAN PROBABILITY DISTRIBUTION STANDARD DEVIATION = 2 = X ~
15
GAUSSIAN (NORMAL) PROBABILITY DISTRIBUTION
16
ASSUMPTIONS FOR A NORMAL PROBABILITY DISTRIBUTION SAMPLE SELECTED FROM A LARGE POPULATION SAMPLE = HOMOGENEOUS STOCHASTIC = RANDOM MEASUREMENT PROCESS NO SYSTEMATIC ERRORS AFFECTING THE RESULTS
17
GAUSSIAN (NORMAL) STATISTICAL DISTRIBUTIONS MEAN - 1 STD < X < MEAN + 1 STD –CONTAINS 68.3 % OF MEASUREMENTS MEAN - 2 STD < X < MEAN + 2 STD –CONTAINS 95.5 % OF MEASURMENTS MEAN - 3 STD < X < MEAN + 3 STD –CONTAINS 99.7 % OF MEASUREMENTS
18
GAUSSIAN (NORMAL) PROBABILITY DISTRIBUTION
19
Poisson Probability Distribution Function Poisson distribution: m = mean, shape governed by one variable P(x) difficult to calculate for large values of x due to x! X-ray and -ray counting statistics obey P(x) Used to describe Radioactive decay Quantum mottle
20
Probability Distribution Functions Probability of observing an observation in a range: integrate area (for G): 1 σ = 68.25% 1.96 σ = 95% 2.58 σ = 99% Error bars and confidence intervals P(x) very similar to G(x) when σ ≈ √x → use G(x) as approx. Can adjust the noise (σ) in an image by adjusting the mean number of photons used to produce the image c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., pp. 276 - 277.
21
GAUSSIAN (NORMAL) DISTRIBUTION EXP[ - ( X - X ) 2 / 2 2 ] (2 ) 0.5
22
COMPARISON OF VARIOUS STATISTICAL DISTRIBUTIONS OF PROBABILITY FOR COIN FLIPPING
23
Quantum Noise N = mean photons/unit area σ = √N, from P(x) → σ 2 (variance) = N Relative noise = coefficient of variation = σ/N = 1/√N (↓ with ↑ N) SNR = signal/noise = N/σ = N/√N = √N (↑ with ↑ N) Trade-off between SNR and radiation dose: SNR ↑ 2x → Dose ↑ 4x c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 278.
25
Noise Frequency: the Wiener Spectrum W(f) Although noise appears random, the noise has a frequency distribution Example: ocean waves Take a flat-field x-ray image (still has noise variations) Fourier Transform (FT) the flat image → Noise Power Spectrum: NPS(f) NPS(f) is the noise variance (σ 2 ) of the image expressed as a function of spatial freq. (f) c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 282.
26
Detective Quantum Efficiency DQE: metric describing overall system SNR performance and dose efficiency DQE = SNR 2 in = N(→ SNR = √N) SNR 2 out = DQE(f) = c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 282. DQE(f=0) = QDE
27
Contrast Detail (C-D) Curves Spatial resolution: MTF(f) Contrast resolution: SNR Combined quantitative: DQE(f) Qualitative: C-D curve C-D phantom: holes in plastic of ↓ depth and diameter What depth hole at which diameter can just be visualized Connect the dots → C-D line Better spatial resolution: high- contrast, small detail Better contrast resolution: low- contrast c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 287.
28
Receiver Operating Characteristic Curves The ROC curve is essentially a way of analyzing the SNR associated with a specific diagnostic task A z : area under the curve – concise description of the diagnostic performance of the systems (including observers) being tested Measure of detectability A z = 0.5 guessing A z = 1.0 perfect c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p. 291.
29
Receiver Operating Characteristic Curves Diagnostic task: separate abnormal from normal Usually significant overlap in histograms Decision criterion or threshold Based on threshold: either normal (L) or abnormal (R) N cases: 2 x 2 decision matrix TPF= TP/(TP+FN)= Sensitivity FPF = FP/(FP+TN) Specificity = (1-FPF) = TNF ROC curve: sensitivity vs. 1- specificity usu. @ five threshold levels c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., pp. 288-289.
30
ROC Questionnaire: 5 Point Confidence Scale
31
Rank Signal (Lesion) Detection On A Scale of 1 to 5. 1. Almost certainly NOT present. 2. Probably NOT present 3. Equally likely to be Present or Not Present. 4. Probably PRESENT 5. Almost certainly PRESENT Make a table of the number of cases receiving each rank for both the positive and negative images. The ROC Cookbook
32
Categories Image Rank (Certainty that a lesions is present) 1 Certainly Not 2 Probably Not 3 Unsure 4 Probably Present 5 Certainly Present Total Number of Images Positive Images 21434 16100 Negative Images 2451 213150 The survey
33
Make a second table with a cumulative ranking: Add the cells so that the lowest rank has the total of all possibilities, the next has all but the lowest rank, the next all but the two lowest rank, etc. Cumulative Rank 1+2+3+4+52+3+4+53+4+54+55 Positive Images 100 98 84 50 16 Negative Images 15012675243 Make the Cumulative Table
34
Divide the positive image values by 100 Divide the negative image values by 150. Put them in a new table. Cumulative Rank 1+2+3+4+52+3+4+53+4+54+55 Positive Images 1.98.84.50.16 Probability of calling a signal when a signal is present. Negative Images 1.84.5.16.02Probability of calling a signal when a signal is absent. Normalize the Data to One.
35
Plot the results. The straight line is a pure guess line. The area under the curve is Az, a measure of overall image performance. Az = 0.5 is equivalent to pure guessing. The greater the area under the curve, the better the system under test performs the task. Plot the Curve